Product Manual

How-To Use Assure® ID Insulin Safety Syringes
Step 1
To open, separate paper backing from plastic by grasping tabs and peeling apart with both hands.
Step 2
Hold onto safety shield while removing protective needle cap. Fill syringe with insulin while
holding index finger against finger flange.
Step 3
Extend safety shield forward until you feel the positive stop and hear the audible snap. This will
secure the safety shield for safe transportation to patient site. Do not twist safety shield or
syringe. Once shield is locked by twisting, syringe should be discarded.
Step 4
To give injection, keep fingers clear of open end of safety shield, retract safety shield until it
snaps into original position. For safety reasons, it will take a little extra force to retract safety
shield to original position. To avoid unexpected extension of the safety shield, always keep one
finger over the finger flange during use.
Step 5
Immediately following injection, extend safety shield forward until snap is heard. Then, lock
safety shield by twisting in either direction until you feel the positive stop and hear an audible
click.