13.6 Initial
moisture an
a
l
ys
i
s
To determine optimal drying parameters for unknown sample, it is recommended to
perform initial measurement with activated drying chart displaying. To do this, set the
following drying parameters (see Drying parameters setting):
- Operation Mode: Time mode
- Calculation method: m0-m/m0*100%
- Drying temperature:
organic substances: 80 - 120
0
C
inorganic substances: 140 - 160
0
C
- Samples quantity: do not set
- Sampling interval: 1 second
- Drying time: set time, after which the sample will be definitely dried
More information regarding temperature and drying time in A appendix.
To activate displaying of drying chart, which will be visible on the display instead of
humidity indication, perform the following actions:
USER MENU
1. Drying settings
2. Memory settings
3. Drying
options
4. Drying raport
5. Configuration
6. Settings default
7. Exit
DRYING OPTIONS
1. Average
2. Drying
chart
3. Transmission
4. Exit
Tare the moisture analyzer with empty single-use pan (T key).
Put a sample of examinated material on single-use pan, put it into drying chamber and
press START key.