User manual

USER MANUAL
39
T
r
o
u
bles
h
ooti
n
g
:
Display indication
Possible cause
Remedy
Initial heating Td
temperature exceeds
105°C, the sensor
does not react when
touched with a finger
The temperature sensor is
damaged
.
Contact an authorised service
poin
t.
Initial heating Td
temperature does
not reach 105°C
during 3 minutes
time, the halogen
heater(s) do not light.
The heater is damaged.
Replace the heater.
"Test ..."
Auto-tests in progress /
electronic unit damage
wait for 1 minute
" - - - - "
The moisture analyzer is
during zeroing
/ mechanical damage
wait for 1 minu
t
e
check if the moisture analyzer
is placed on stable ground, no
t
affected by vibra
t
ions
Tare range
exceeded
Tare key pressed during zero
indica
t
ion
Moisture analyzer indica
t
ions
must be different than zero
Zeroing range
exceeded
Permissible zeroing range
was exceeded
Remove the load from the pan
Weighing range
exceeded
Permissible weighing range
(Max +9e) was exceeded
Reduce the load
„Measuring range
exceeded (+)
Upper limit of analog-digi
t
al
transducer measuring range
was exceeded
Remove the load from the pan
„Measuring range
exceeded (-)
Lower limit of analog-digi
t
al
transducer measuring range
was exceeded
Check if there are all necessary
pan elemen
t
s