User`s manual

29
“Enter” key to confirm.
3.1.4 34.5V Calibration
Press “Enter” key to verify and then the display will show as follows:
Input the value you read from KEITHLEY 2000, which is retained to four decimal
places. Press “Clear” key to remove the wrong input value of current digit. Then press
“Enter” key to confirm and exit to calibration menu. Now, voltage calibration is
completed.
3.2 CC Mode Calibration
In this mode, three current points: 0.5A6A9A should be calibrated.
3.2.1 Wiring
-
+
OUTPUT
366XA
FLUKE45
+10A
-
3.2.2 0.5A Calibration
Press “Enter” key, you will see the following figure in the display: