Specifications
E-500/E-501 Series PZT Servo Controllers User Manual PZ 62E
Release 2.15.1 www.pi.ws Page 33
Labeling Type Function
CONTROL INPUT
-2 to +12 V
BNC In analog operation, this control input voltage gives the target (either
as voltage or position, depending on the servo mode). The input
signal should always be in the range of 0 to 10 V (excursions to -2 or
+12 V may cause overflow, especially with servo on, and reduce
actuator lifetime).
The control input range can be shifted using the “DC-OFFSET”
potentiometer.
The control input voltage can also be a computer-generated analog
signal (e.g. from a DAQ board). You can use a PI LabVIEW Analog
Driver set to generate that analog signal. See “Analog Operation” on
p. 17 for details.
TEMP SENSOR
LEMO
EPL.0S.303.HLN
Connection for PT1000 temperature sensor or dummy plug.
Pinout on p. 35
PZT
-30 to +130 V
LEMO
EGG.0B.302.CLL
Pinout on p. 35
Output of the piezo voltage for the piezo actuator in the stage.
Piezo voltage output is deactivated if a temperature of 150 °C is
exceeded at the piezo actuator. Automatic reactivation at a
temperature < 146 °C