Specifications

E-500/E-501 Series PZT Servo Controllers User Manual PZ 62E
Release 2.15.1 www.pi.ws Page 41
For example, in open-loop operation, the output range +3 to +1100 V and gain +100 yield a
Control In range of 0 to +11 V.
Setting the Output Range Jumper
Ranges
E
+3 to +1100 V
(factory setting)
D
-260 to +780 V
C
-550 to +550 V
B
-780 to +260 V
A
-1100 to -3 V
To change setting, remove red cap
(shown in Range E position) and
replace over pins corresponding to
desired range.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com