Specifications
E-500/E-501 Series PZT Servo Controllers User Manual PZ 62E
Release 2.15.1 www.pi.ws Page 44
Labeling Type Function
With E-509.E3 ony:
OFL
LED
Yellow/off
Overflow state:
Yellow: Overflow condition, i.e. the amplifier is near its range
limit.
Off: No overflow condition
When the OFL LED comes on, a zero-point adjustment of the
sensor can be necessary. See p. 17 and the E-509 User Manual
(PZ77E) for details.
ZERO
Trim
potentiometer
A trimmer adjustment tool can be used on the ZERO potentiometer
for a zero-point adjustment of the sensor. A zero-point adjustment
can be necessary after longer operation (changes in temperature)
or if the load on the piezo stage is changed.
SENSOR MONITOR
LEMO
EGG.0B.306.CLL
Output of the monitor signals for the sensor channels.
Pinout on p. 48.
5.8.3. Front Panel Elements of Modules for Strain Gauge Sensors
Fig. 24: E-509.S1 sensor / servo
controller
Fig. 25: E-509.S3 and E-509.S31 sensor /
servo controller
INFORMATION
In the labeling of E-509.S1 and E-509.S3 modules, X stands for S.
INFORMATION
E-509.S31 and E-509.S3N models for use with P-611 NanoCube® stages:
E-509.S31 differs from E-509.S3 only in the settings of the internal E-801 sensor submodule.
E-509.S3N consists of an E-509.S31 module plus a P-611.91 3-channel adapter cable
LEMO/Sub-D 25 (f).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com