Specifications
System
Overview
Reactor Assembly
Solvent Bottle Assembly
The solvent bottle assembly consists of a
4L
Solvent Bottle, plastic Bottle Safety
Carrier, and a valved Teflon cap. The solvent bottle cap has three Teflon lines attached
to it: One to supply pressure to the bottle, the second to allow venting of the bottle, and
the third is a liquid line which leads to the Quest Reactor Unit. The pressure and
delivery tubes protrude from a point within the umbilicus. The vent line, found in the
accessory pack, simply connects to the bottle cap and should be pointed away from the
user and anchored within the fume hood. On the top of the Solvent bottle cap is a dial.
To pressurize the bottle, rotate the dial so that the arrow is pointed to the Teflon tube
labeled "Pressure." To vent the bottle, rotate the dial so that the arrow is pointed to the
Teflon tube labeled "Vent."
WARNING Always vent the bottle prior to unscrewing the cap.
When the bottle cap dial is in the "pressure" position,
8
psi gas flows from the control
box into the solvent bottle by means of the pressure line.
-
WARNING Always place the
4
L
solvent bottle in the plastic Safety Carrier when
working with the Quest.
-
WARNING Always wear eye protection when operating the Quest.
-
A long Teflon line runs from the bottom of the Teflon cap. This line is the "pickup" tube
for the solvent in the bottle. This line should reach the bottom of the bottle.
-
Additionally, a polyethylene filter should be placed on the end of the line to prevent
particulates from entering the Quest Fluidic System.
-
When installing a
4L
bottle on the Quest, ensure that all
4
Manifold Control Valves are
set to "Closed."
-
To install a
4L
bottle, simply screw the outside part of the bottle cap down on the neck
of the bottle. Ensure that an new end-line filter is installed on the solvent line and a
Teflon bottle seal is present in the cap. Turn the dial on the top of the cap to point to the
-
tube marked "Pressure." Wait a minute or so for the bottle to pressurize.