Specifications
Control Box Connections to the Reactor Assembly 
Control Box Connections to the Reactor Assembly 
The control box has connections for the Umbilicus for interface to the Reactor Unit. 
These ports are: 
Heater Power 
- 
Supplies power to the heaters within each reaction 
vessel bank. 
Signal 
- 
This cable carries thermocouple readings from the Reactor 
Unit to the Controller Unit. 
Agitation Gas 
- 
Two ports on the Controller unit supply 
HP 
gas to the 
pneumatic gas cylinder. 
Solvent Gas Pressure 
- 
A 
port on the Controller Unit is used to supply 
gas pressure to the Solvent Bottle Assembly. 
Drain Gas 
- 
The drain gas port supplies 30 psi gas line from the 
Contoller Unit to the Reactor Unit. 
Metered Gas 
- 
The drain gas port supplies 10 psi gas line from the 
Contoller Unit to the Reactor Unit, and is flow regulated by a needle 
valve on the Reactor Unit. 
Controller Unit Gas and Power Connections 
System Gas 
The gas port (located on the rear of the Controller Unit) uses Quick Connect fittings to 
interface with a inert house gas system. A minimum of 30 psi is required. (Maximum 
pressure 
= 
40 psi) 
Electrical Outlet 
The power cord connects to the rear of the Quest Controller Unit. The instrument can 
operate on 90-240 VAC wall outlets. 
Power Switch 
After you have plugged in the power cord to the power outlet, the power switch turns 
the power on or off. 