Specifications
Installing New Reaction Vessels
I
---
Set
Temperature
-
-
A:
90C
1:00
OFF
B
:
60C
0.00
OFF
RVSize
5mL 5mL
b.
Press the Forward button to move the cursor to the RV size line.
c.
Press either one of the Parameter buttons to set either
5
ml
,
or 10 ml vessel
sizes.
2.
If the reaction is solid phase, weigh out and add the synthesis support, or add
ArgoCaps, to the reaction vessels. The maximum amount of resin that can be used
depends on the resin density, resin swelling characteristics, and solvents used for
synthesis. As a general rule, up to 200 mg of
ArgoGel, or similar gel type resin, can
be used in the
5
ml
vessels.
3.
With a soft cloth and appropriate solvent, clean the male reaction vessel fittings on
the Upper and Lower Manifolds. Ensure that there are no particulates or dust on the
fittings.
4.
Place magnets into the reaction vessels.
5.
Place one of the Reaction Vessel Tubes in the appropriate channel and press down
to the top of the lower manifold male fitting.
Do
not attempt to fully seat the
reaction vessel at this step.
6.
Place the Reaction Vessel Insertion Tool on the top of the reaction vessel, with the
nub of the tool inserted into the vessel.
QUEST
210
User
Manual