Specifications
Operating the Quest 
210 
Starting a Reaction 
Temperature Control 
- 
The allowable reaction temperature range on the Quest is 
-40 
to 
+130. 
For a particular 
reaction, do not program a temperature set-point which exceeds the boiling point of the 
reaction solvent. 
Each side of the Quest may have a different set point. To achieve sub-ambient 
temperatures, an external recirculating chiller is required. When using the chiller, only 
connect it to one side of the Quest at a time. 
WARNING  When heating, the Reactor 
Assembly will become hot. The symbol at the 
right is to warn of hot surface. Use caution when 
working near or on the Reactor Assembly. 
Allow it to cool before attempting to access RVs. 
Prior to starting a heating step, set both Manifold Control Valves to "Closed".  This 
prevents solvent loss during heating: 
drain 
solvent 
G 
closed 
utility 
1 
metered gas 
vent 
0 
closed 
utility 
2 