Specifications
Operating the Quest 
210 
Starting 
a 
Reaction 
9. 
Make sure that the vessel size selection in the Set Temperature screen matches the 
RV 
size that is being used for the synthesis. If not modify the selection in the screen. 
10. 
The temperature set point can be changed from an existing set point, by using the 
Set Temperature screen. 
11. 
The heaters will turn off when the time has elapsed, or they can be stopped 
immediately by pressing the 
STARTISTOP 
button with the cursor on the 
LCD 
line 
corresponding to the 
RV 
bank. 
12. 
If 
desired, press Mode button to view the Monitor screen. This screen will display 
the set point and the actual 
RV 
temperature. 
Chilling the Reaction Vessels 
For sub-ambient temperature set points, an external recirculating chiller is required. 
Set the low temperature set-point into the Controller Unit in the same manner as was 
described previously for heating. 
Use the chiller on only one side of the Quest. 
If 
a second chiller is available, then both 
sides of the Quest can be chilled together. 
Connect the Chiller Fitting Kit Accesory to the Quest 
Figure 
3-8 
Quest 
Chiller 
Fitting 
Accesory 
a.Using  a 
3/32" 
hex wrench, remove the 
4 
screws fastening the heater block guard 
to the front of the heater assembly. 