Specifications

Temperature
Control
-
b.Wrap teflon tape around the threads of both of the 1/16" elbow fittings exiting the
cooling channel on both sides of the heater assembly.
c.Using a 314" socket (either
an
open end wrench or crescent wrench may also be
used), tighten the chiller accessory fitting assembly to the 1/16" elbow fitting
exiting the heater assembly.
d.Push on the
2
Y2"
piece of 318" thick pipe insulation provided.
e.Instal1 the the heater block guard (pin 101634) to the front of the heater assembly.
f.Repeat for the other heater if desired.
Connect the chiller outlet hose to the 114
"
female pipe brass fitting at the left of the
reaction vessel bank. Ensure that the chiller hoses and brass fittings are equipped with
foam insulation.
Connect the chiller return hose to the 114
"
female pipe brass fitting on the right of the
reaction vessel bank. Ensure that the chiller hoses and Brass fittings are equipped with
foam insulation. (Insulated recirculating chiller tubing can be purchased from Julabo).
Turn on the chiller to flow the coolant liquid through the Quest cooling channel.
On the Quest Controller Unit, press the Mode button, until the "Monitor" screen is
displayed.
Adjust the bath temperature of the chiller until the Actual
RV
temperature matches the
Set Point for the reaction vessel bank.
- --
WARNING
Disconnect the chiller when not in use. Purge the cooling channel of
refrigerant liquid after a the chiller is disconnected. Do not heat a reaction vessel bank
with the chiller connected, or with residual liquid in the cooling channel.
QUEST
210
User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com