Parallel Solvent Addition and Vessel Draining 
Draining the Reaction Vessels 
Reaction Vessel draining is typically done during solid phase synthesis during resin 
wash steps, or when a reaction step is complete. 
1. 
Verify that the rectangular glass waste container is installed below the reactor 
assembly. 
2. 
Toggle the "Upper Manifold Membrane" switch to "Open." 
3. 
Set the Manifold Control Valves as follows: 
drain 
solvent  closed 
utility 
1 
metered gas 
vent 
0 
closed 
utility 
2 
4. 
Open the drain valves on the lower manifold 
To open the valves individually: 
Push down on the drain valve lever to drain the 
reaction vessel. Return the lever to the closed (up) position when finished. You can 
open multiple valves at the same time. 
To open ten valves at the same time: 
Use the parallel drain lever, which fits over 
all 10 drain levers, to control all 10 valves simultaneously. 
5. 
After draining the vessels, set both Manifold Control Valves to "Closed: 
drain 
solvent  closed 
utility 
1 
QUEST 
21 
0 
User 
Manual 
metered gas 
vent 
0 
closed 
utility 
2 