Specifications
Collection of Liquid Reaction Products 
1. 
Toggle the "Upper Manifold Membrane" switch to "Open." 
2. 
Prior to product collection, it is recommended that the system be vented to relieve 
any residual pressure that can cause splashing during product collection. To vent the 
vessels, rotate the Manifold Control Valves as shown below. Slowly approach the 
"Vent"  setting on the valve to vent the pressure slowly to avoiding "bumping" the 
solution. 
drain 
solvent 
n 
closed 
utility 
1 
metered gas 
vent 
0 
closed 
utility 
2 
3. 
Carefully lower and remove the Glass Collection Tank, and drain its contents to an 
appropriate storage container. 
4. 
Place the desired collection rack and vials on the waste collection tray. 
5. 
Set the Manifold Control Valves as follows to pressurize the head space over the 
reaction vessels: 
drain 
solvent  closed 
utility 
1 
metered 
gas 
vent 
0 
closed 
u 
utility 
2 
6. 
Open each drain valve on the lower manifold (down lever position) one at a time. 
The solution should drain into the collection vial. Return the valve to closed (up 
position) after each collection. The flow rate of the Metered gas may need to be 
adjusted for the this step 