Specifications
Revision
06/25/98
Argonaut
Technologies
A. Attach the single end of the Gas Supply Assembly
(Part
Number 300453) to an inert gas
source capable of providing a regulated output between 55-60 PSI. The other end of the
Gas Supply Assembly has a tee with two quick disconnect fittings attached. Connect one
quick disconnect fitting to the inert gas inlet and the other fitting to the compressed air
inlet on the rear of the ASW Controller Unit. To attach these fittings push them over the
appropriate inlet port until an audible click is heard. Continue the installation at step
3.
Note:
If
either gas source is greater than five feet from the
ASW
Controller Unit additional gas
supply tubing will be required for dual gas operation.
B.
Remove the tee and quick disconnect fittings from the end of the Gas Supply Assembly
(Part Number 300453). Cut the remaining 10
ft
piece of tubing in half. Install one quick
disconnect fitting onto the end of each piece of tubing to create two inlet lines. Attach the
open end of the first inlet line to an inert gas source capable of providing a regulated
output between 40-60 PSI. Attach the quick disconnect fitting on the opposite end to the
inert gas inlet on the rear panel of the ASW Controller Unit. Attach the open end of the
second inlet line to a compressed air source capable of providing
a
regulated output
between 55-60PSI. Attach the quick disconnect fitting on the opposite end to the
compressed air inlet on the rear panel of the ASW Controller Unit. To attach these
connectors push them over the appropriate inlet port until an audible click is heard.
Figure
2
-
ASW
Controller Unit Rear View
3.
Refer to Figure
2
for the location of components on the rear panel of the ASW Controller Unit.
Ensure the onioff switch is set to OFF and connect the power cable to the electrical socket on the
rear of the Controller Unit. Connect the free end of the power cable to an appropriate power
outlet as specified in the Quest User Manual.
4. See Figure
3
for the location of the ASWM interface harness connection on the front of the ASW
Controller Unit. Attach the ASWM Interface harness to the fiont of the Controller Unit by lining
up the surfaces and rotating the locking collar in a clockwise direction until tight.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com