Specifications

Revision
06/24/98
Argonaut Technologies
Note:
If
this Automated Solvent Wash Option is being installed as part of an initial Quest
install refer to the Quest User Manual for additional information concerning setup of the
Reactor Unit.
9.
If the Automated Solvent Wash Option was received as an upgrade to an existing Quest continue
with step 10, otherwise proceed with step 11.
10.
Three Bottle Cap Assemblies are included when the Automated Solvent Wash Option is shipped
as an upgrade to an existing Quest 210. Use the Bottle Cap Assembly originally shipped with the
Quest as the fourth assembly. To connect the fourth Bottle Cap Assembly locate the solvent
tubing marked DS and the pressure tubing marked DP in the ASWM. Connect the tubing to the
corresponding ports on the fourth bottle cap assembly per Figure
5.
o+
Pressure
tubing
I
0
1
Solvent tubing
Figure
5
-
Bottle Cap Assembly
Note: The tubing attached to the Bottle Cap assemblies in the ASWM are labeled with two
characters. The first character, either A,
B,
C,
or
D
describes the bottle cap assembly it
attaches to. The second Character, either
P
or S describes whether it is solvent or pressure
tubing.
Note: to allow for flexibility in locating the solvent bottle vents, additional tubing assemblies
(part number
300319)
are provided for connection to the Vent ports on the Bottle Cap
Assemblies.
1
1.
Place four solvent bottles in the safety carriers and remove the clear cap from the top of the each
carrier. Place each solvent bottle with carrier into the ASWM. ~ndividuall~ remove the solvent
bottle caps, insert a Teflon bottle seal and attach a Bottle Cap Assembly. The Bottle Cap
Assemblies are labeled A, B,
C,
and D. Table
2
identifies where in the ASWM each of the
solvent bottles should be placed and which Bottle Cap Assembly should be attached. If installing
solvents bottles the first time, refer to "Manual Operation Verification" for the recommended
solvents to use for that orocedure.
1--
-
~
Bottle Cap Assembly
I
Location in the ASWM
(as viewed from the front)
A
- -
B
C
Table
2
-
ASWM Solvent Bottle Locations
Left
Front
-
-
-
.
-
.
-
. .
.
Right Front
Left Rear
I
-
-
-
.
-
-
.
.
.
.
Note: For all customers outside of North America solvent bottles are provided.
D
Q~JEsT.?
10
ORGANIC
SYNTHESIZER
-
A
UTOW TED
SOISJIINT
WASH
Right Rear