Specifications
Argonaut Technologies  Revision 
06/24/98 
12. 
Using Figure 
5 
as a guide, turn the valves on the top of the Bottle Cap Assemblies towards the 
vent port.  This completes the installation of the Automated Solvent Wash Option.  Before using 
the Automated Solvent Wash Option it is recommended that you perform the Automated Solvent 
Wash Operational Checkout. 
WARNING:  Always vent the bottle prior to unscrewing the Bottle Cap Assembly. 
WARNING:  Always place the solvent bottles in the plastic Safety Carrier when working with 
the Quest. 
Once the Automated Solvent Wash Option installation is complete, performing the Operational 
Checkout will insure that the system has been properly installed and functions correctly. The first 
procedure verifies the configuration of the ASW Controller Unit. The second procedure will verify 
manual operation of the Automated Solvent Wash Option by demonstrating that a specific bottle will 
deliver solvent when selected. 
WARNING 
- 
Always wear eye protection and appropriate clothing when operating the Quest. 
7.1 
Auto Detection of Automated Solvent Wash Configuration 
1. 
Turn ON the ASW Controller Unit and the gas supply. 
2. 
On power up, the Quest firmware will determine ifthe hardware and software necessary for Auto 
Solvent Wash are present. 
3. 
Pressing the Mode Key, located on the front of the Controller Unit, once will display the 
Firmware revision screen. 
If 
the firmware version screen includes the line "with  Auto Solv. 
Wash"  as shown in Figure 
6, 
then the Controller Unit is properly configured. 
QUEST 210 
----- 
with Auto Solv. Wash 
Date:  Jun 12, 
1998 
Figure 
6 
- 
Firmware Version Screen 
Note:  The firmware revision screen can also be accessed from the OTHER FUNCTIONS menu. 
7.2 
Manual Operation Verification 
Note: 
It 
is recommended that methanol be used as the solvent when performing the Manual 
Operation Verification procedure. 
1. 
See Figure 
4 
for the location of the toggle valves on the ASWM. Verify that the 
5 
toggle valves 
on the ASWM front panel are all in the off (down) position. 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com










