Specifications

Argonaut Technologies Revision
06/24/98
12.
Using Figure
5
as a guide, turn the valves on the top of the Bottle Cap Assemblies towards the
vent port. This completes the installation of the Automated Solvent Wash Option. Before using
the Automated Solvent Wash Option it is recommended that you perform the Automated Solvent
Wash Operational Checkout.
WARNING: Always vent the bottle prior to unscrewing the Bottle Cap Assembly.
WARNING: Always place the solvent bottles in the plastic Safety Carrier when working with
the Quest.
Once the Automated Solvent Wash Option installation is complete, performing the Operational
Checkout will insure that the system has been properly installed and functions correctly. The first
procedure verifies the configuration of the ASW Controller Unit. The second procedure will verify
manual operation of the Automated Solvent Wash Option by demonstrating that a specific bottle will
deliver solvent when selected.
WARNING
-
Always wear eye protection and appropriate clothing when operating the Quest.
7.1
Auto Detection of Automated Solvent Wash Configuration
1.
Turn ON the ASW Controller Unit and the gas supply.
2.
On power up, the Quest firmware will determine ifthe hardware and software necessary for Auto
Solvent Wash are present.
3.
Pressing the Mode Key, located on the front of the Controller Unit, once will display the
Firmware revision screen.
If
the firmware version screen includes the line "with Auto Solv.
Wash" as shown in Figure
6,
then the Controller Unit is properly configured.
QUEST 210
-----
with Auto Solv. Wash
Date: Jun 12,
1998
Figure
6
-
Firmware Version Screen
Note: The firmware revision screen can also be accessed from the OTHER FUNCTIONS menu.
7.2
Manual Operation Verification
Note:
It
is recommended that methanol be used as the solvent when performing the Manual
Operation Verification procedure.
1.
See Figure
4
for the location of the toggle valves on the ASWM. Verify that the
5
toggle valves
on the ASWM front panel are all in the off (down) position.