Specifications
2.
Close the Upper Manifold by first turning the
left
Manifold Control Valve to
"CLOSED", then turn the right Manifold Control Valve to "CLOSED".
Drain Gas Metered Gas
Utility
1
Utility
2
D.
Adjusting the Agitation Bar
1.
Ensure that the Agitator Bar is in the
UP
position (move the agitator up by pushing
the Agitation UP button located to the right of the Quest Controller Unit display).
2.
Use the
7/64"
hex wrench supplied with the Quest to loosen the Agitation Stops.
Adjust the Agitation Stops on the Reactor Unit so the maximum magnet height is
-2
cm below the solvent level in the RVs. Tighten the Agitation Stops so both stops are
at the same height (see
Figure 2).
3.
On the Quest Controller Unit press the MODE soft key located below the LCD
display until the Agitation menu is displayed
(Figure 12).
----
-----
-
--------
MixEvery
:
4.0
sec
Upstroke
:
3.8
sec
%
Upward
:
95
%
-
-
Not Agitating
-
-
-
Figure 12. Agitation LCD Display
E.
Agitation Cycles for ArgoCap Washing
The following washing/agitation regimen is recommended for use with polystyrene resin
encapsulated ArgoCaps. The agitation is gradated in order to provide gentle agitation for
the initial dissolution of the ArgoCaps. Once the ArgoCaps have dissolved and the resin
is swollen in DCM solvent, the agitation is set for a gentle cycle to be used throughout
the synthesis. Please note that these Parameter Settings are representative and may vary
somewhat from instrument to instrument.
1.
Press the Right or Left soft keys
(",)
to select the "Mix Every" parameter (agitation
frequency) on the Agitation menu.
2.
Use the PARAMETER SETTING soft keys
(?&)
to increase or decrease the agitation
frequency to the following: Mix Every
4.0
sec.
Argonaut Technologies
Part
Number 102298
04/02/98
Quest Tutorial