Specifications
GRCM 
INSTALLATION 
' 
Installation Steps 
, 
3. 
Attach the end of that line to side 
B 
of the 
"T" 
(Part number 101955) and 
make it finger-tight. 
4. 
lnsert one end of the lnert Gas Extension Line (Part number 300797) into 
side 
A 
of the 
"T" 
and make finger-tight. lnsert the other end of the line into 
port #3 of the Quest controller and make finger-tight. 
5. 
Attach one end of the lnert Gas Inlet Line (Part number 300795) to side 
C 
of 
the 
"T". 
Connect the other end to input 1 of the GRCM Control box. See 
Figure 3 below. 
6. 
lnsert the 118  inch PEEK fitting of the reagent gas inlet line (Part number 
300796) to input 
#2 
of the Control box and make finger-tight. Connect the 
compression fitting on the other end of this Teflon tubing to the outlet of 
the reagent gas regulator. Do not set the regulator above 30 psig (3.08 
barr). 
See Figure 3 below. 
Figure 
3: 
Control Box showing Gas Connections 
1 
and 
2 
GRCM 
- 
User 
Manual 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com










