Specifications
Operation 
The following list shows the step-by-step procedure for 
carrying out concentration  by  evaporation with the setup shown above. 
1.  Set up the Quest and GRCM as shown in Figure 
9 
2.  Connect the Solvent Trap Manifold to a solvent trap. 
3. 
Turn each of the two Quest Control valves to CLOSED. 
4. 
Turn the Gas Selection valve on the GRCM Control Box to Gas 
1 
(Nitrogen). 
5. 
Slowly turn the Out 
A 
needle valve counterclockwise approximately 
% 
turn 
to deliver nitrogen. 
/7\ 
Caution 
Excessive opening 
of 
the Out 
A 
needle valve at this time may 
cause the liquids in the 
RV 
to "bump" (bubble excessively). 
6. 
Carefully open the Quest Lower Manifold. 
7. 
Slowly turn the Out 
A 
needle valve counterclockwise to adjust the gas flow. 
To  achieve fastest evaporation rates, maximize the gas flow without 
allowing the liquid to bubble over.  As the liquid level decreases, the gas 
flow can be readjusted to maintain maximum evaporation rates. 
8. 
If necessary, apply mild heating (approx. 
213 
of boiling point) for efficient 
evaporation. 
9. 
When the solvent is evaporated, close the Lower Manifold drain levers on 
the Quest reactor. 
10. Close the Out 
A 
needle valve by turning it clockwise until it stops. 
11. Remove the Solvent Trap and Gas Delivery Manifolds from the Quest 
reactor. 
12. Clean the apparatus as decribed in the section "Cleaning the GRCM" 
below. 
GRCM 
- 
User 
Manual 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com










