Installation Guide
teleCARE IP
5.6.1 Auxiliary Lamp Connection - NILD2
The green LED board (NILD2-GAA) has a galvanically separated output that can be used to connect to
an auxiliary lamp. It has a maximum switching capacity of 0.4A at 60V peak.
Figure 48. External corridor lamp connections through the NILD2-GAA
IMPORTANT: In order to retain the galvanic separation, it is not allowed to connect the GND
(-) of the auxiliary power supply to the GND (-) of the NIRC3.
5.6.2 Connecting the Wires to the NILD2-GAA
The 2-pole connector terminal on the green LED lamp board (NILD2-GAA) has two terminals with one
connection point at each terminal. A connection point accepts one solid wire of maximum wire size
0.5MM² (0.8mmØ). To connect the wires first strip 6 mm of the insulation from the end of each wire
which is to be connected.
NILD2
To auxiliary power supply
Max. 0.4A at 60V peak