User's Manual

TD 93021US Rev A
10 FEB 2014 / Ver A 33
Installation Guide
teleCARE IP
Four wires are required for the room bus, passive bus and light relay outputs, so repeat the above
illustrated procedure on the remaining three wires.
Figure 30. Connector terminal complete with four wires
Disconnecting a Wire from the Connector Terminal
First carefully place the point of a small screw driver (point approximately 0.1in (2.5mm) wide) on the
relevant orange colored release key of the connection terminals and press the key in firmly to open the
spring-cage connector (1). With the release key pressed in pull the wire from the terminal (2) then
remove the screw driver.
Figure 31. Removing a wire from a connection point
5.4.2 2-Pole Connector Terminal (NICT-2BA)
Figure 32. 2-pole connector terminal