User's Manual

TD 93021US Rev A
10 FEB 2014 / Ver A 66
Installation Guide
teleCARE IP
The designation of the passive bus wires is shown in the following illustration.
Figure 75. 4-pole connector terminal with the passive bus
The designation of the light switching relay connections is shown in the following illustration.
Figure 76. 4-pole connector terminal for the light relay
6.4.2 Preparing the Wires for the 4-pole Connector Terminal
The 4-pole connector terminal has four terminals with two connection points at each terminal. Each
connection point accepts one solid wire of maximum wire size 24 gauge (0.5mm²) 0.03inØ (0.8mmØ).
To connect the wires first strip the jacket from the cables and pull the wires through the housing
leaving a length of 6in (150mm) free, as described in section 5.3, Preparing the Room Bus and
Power Cables on page 26. Then strip 0.25in (6.5mm) of the insulation from the end of each wire
which is to be connected.
Figure 77. Stripping the wires for insertion in the connection terminal
6.4.3 Connecting the wire in the 4-pole Connector Terminal
After stripping the wire to expose 0.25in (6.5mm) of conductor, insert the wire in the appropriate
opening of the connection point by pressing the wire firmly into the terminal, as illustrated below.
Note: Each connection point in the connector terminal accepts only one wire. Maximum
wire size 24 gauge (0.5mm²) (0.8mmØ).
5.5Vdc
IN0
IN1
OUT
L1}
L2}