User's Manual

Table Of Contents
TD 93021US
17 July 2017 / Ver. PF3 96
Installation Guide
teleCARE IP
6.13.2 Auxiliary Inputs
The third 4-pole connector is used for auxiliary inputs. It has two input circuits which are available as
interfaces to external devices. These terminals provide 5.5Vdc at 0.6mA, for monitoring the external
device contacts, as shown in Figure 130.
Figure 130. Auxiliary Inputs
WARNING:
The equipment that is connected to this interface is not considered to be part of the system
configuration unless the equipment complies with ANSI/UL 2560 standard for Emergency Call
Systems for Assisted Living and Independent Living Facilities.
The Auxiliary Monitoring circuits are not electrically isolated.
The cable length for the Auxiliary Monitoring circuits must not exceed 33ft (10m) and the cables
must be appropriately separated from power cables and sources of electromagnetic interference.
The Auxiliary Monitoring circuits are power limited to 5.5Vdc at 0.6mA.
Test
LED