User's Manual
Table Of Contents
- Contents
- 1 Introduction
- 2 Practical Engineering Parameters
- 3 System LAN Cabling
- 4 System Power
- 5 Control Equipment
- 5.1 Preparation
- 5.2 Room Controller (NIRC3)
- 5.3 Preparing the Room Bus and Power Cables
- 5.4 Voice Piggy Back (NIVP)
- 5.5 Connection Terminals
- 5.6 Connecting the IP Room Controller Printed Circuit Board
- 5.7 LED Lamp Boards (NILD2)
- 5.8 Corridor Lamp (NICL2)
- 5.8.1 Installing the Corridor Lamp
- 5.8.2 Removing the Corridor Lamp Printed Circuit Board
- 5.8.3 Corridor Lamp Housing
- 5.8.4 NICL2 - Corridor Lamp Electrical Connections
- 5.8.5 4-Pole Connector Terminal
- 5.8.6 Connecting the Corridor Lamp Printed Circuit Board
- 5.8.7 LED Lamp Boards for the Corridor Lamp
- 5.8.8 Connecting the LED Lamp Boards
- 5.8.9 External Corridor Lamp Inputs
- 5.9 System Manager (NISM2)
- 6 Peripherals
- 6.1 Preparation
- 6.2 Installation Instructions
- 6.3 Backplates and Surface Mounting Spacer
- 6.4 Switch Module Electrical Connections
- 6.4.1 4-Pole Connector Terminal
- 6.4.2 Preparing the Wires for the 4-pole Connector Terminal
- 6.4.3 Connecting the wire in the 4-pole Connector Terminal
- 6.4.4 4-Pole Connector Terminal with Looped Wiring
- 6.4.5 Disconnecting a Wire from the Connector Terminal
- 6.4.6 Mounting the Switch Module to the Backplate
- 6.4.7 Mounting the Switch Module to the Surface Mounting Spacer
- 6.4.8 Dismantling the Switch Modules
- 6.4.9 Dismantling a Switch Module from a Spacer
- 6.4.10 Dip Switch Settings
- 6.5 Bedside Module (NIBM2)
- 6.6 Medical Rail Socket (NIMS2)
- 6.7 Door Side Module (NIDM)
- 6.8 Pull Cord Module - Active (NIPC-W3A)
- 6.9 Toilet Cancel Module - Active (NITC-XXA)
- 6.10 Pull Cord Module - Passive (NIPC-XXP)
- 6.11 Toilet Cancel Module - Passive (NITC-XXP)
- 6.12 Pull Cord Module (NIPC2) Active and Passive
- 6.12.1 Mounting the NIPC2 Pull Cord Module
- 6.12.2 Positioning the Back box for the Pull Cord Module
- 6.12.3 NIPC2 Back plate
- 6.12.4 Mounting the NIPC2 Backplate
- 6.12.5 Drilling the Backplate Mounting Holes
- 6.12.6 Mounting the Backplate on the Wall
- 6.12.7 Preparing the Cable for the Pull Cord Module
- 6.12.8 Room Bus Electrical Connections
- 6.12.9 Room Bus Address DIP Switch Settings
- 6.12.10 Passive Pull Cord Module Electrical Connections
- 6.12.11 Mounting the NIPC2 Pull Cord Module to the Backplate
- 6.12.12 Assembling and Attaching the Pull Cord
- 6.13 Duty Selector (NIDS)
- 6.14 Card Reader (NICR)
- 6.15 Speech Module (NISP)
- 6.16 Room Display (NIRD)
- 6.17 Television Interface Module
- 6.18 Sunblind Control Module
- 7 External Inputs
- 8 Wireless Functionality
- 8.1 General
- 8.2 Principle of the teleCARE IP with Wireless Functionality
- 8.3 teleCARE Wireless with Speech
- 8.4 teleCARE IP Wireless Planning
- 8.5 Wireless Infrastructure
- 8.6 Principle of the Wireless Infrastructure
- 8.7 teleCARE IP Wireless Components
- 8.7.1 NUREP Wireless Repeater
- 8.7.2 Outdoor Box
- 8.7.3 Wireless Gateway
- 8.7.4 NUWBM3 Wireless Active Bedside Module
- 8.7.5 NIRX teleCARE IP Transceiver
- 8.7.6 Connecting the Transceiver Module
- 8.7.7 NIVP Voice Piggyback Module
- 8.7.8 NIFX Fixed Transceiver
- 8.7.9 NITX Mobile Transceiver
- 8.7.10 NUUTX Universal Transceiver
- 8.7.11 NUWIR Wireless PIR Module
- 8.7.12 IR Range Test
- 8.7.13 NUUTX NUWIR Battery Placement
- 8.7.14 (3) Slide the battery lid back into place until it snap fits.NUUTX NUWIR Mounting Instructions
- 8.7.15 NILF Low Frequency Beacon
- 8.7.6 NILF Electrical Connections
- 8.7.7 NILF DIP Switch Settings
- 8.7.8 NILF Power Supply
- 9 Installation Examples
- 9.1 2-Bed Room with Active Toilet Cancel and Active Pull Cord Peripherals
- 9.2 2-Bed Room with Passive Toilet Cancel and Passive Pull Cord Peripherals
- 9.3 2-Bed Room with a Medical Rail Socket at each Bed
- 9.4 Room Controller with Corridor Lamps (Master/Slave)
- 9.5 4-Bed Room with Speech
- 9.6 Duty Selector at a staff Station
- 9.7 Positioning of the teleCARE IP Peripherals
- Document History
TD 93021US
17 July 2017 / Ver. PF3 31
Installation Guide
teleCARE IP
5.5 Connection Terminals
The 4-pole and the 2-pole connector terminals required for the room bus and the 24Vdc power supply
are not supplied with the IP room controller. The connectors are available as accessories and must be
ordered separately.
5.5.1 4-Pole Connector Terminal (NICT-4AA)
Figure 28. 4-pole connector terminal
The 4-pole connector terminal is used for connecting the room bus. It has a screw-less “spring-cage”
connection technique and each terminal has two connection points.
The designation of the required four wires is as shown in the following illustration.
Figure 29. 4-pole connector terminal with the room bus
Preparing the Wires for the 4-pole Connector Terminal
The 4-pole connector terminal has four terminals with two connection points at each terminal. Each
connection point accepts one solid wire of maximum wire size 20 AWG (0.8mm).
To connect the wires first strip the jacket from the cables and pull the wires through the housing leaving a
length of 6in (150mm) free, as described in section 5.3, Preparing the Room Bus and Power Cables
on page 25. Then strip 0.25in (6.5mm) of the insulation from the end of each wire which is to be
connected.
Figure 30. Stripping the wires for insertion in the connection terminal
After stripping the wire to expose 0.25in (6.5mm) of conductor, insert the wire in the appropriate
opening of the connection point by pressing the wire firmly into the terminal, as illustrated below.
Note: Each connection point in the connector terminal accepts only one wire. Maximum
wire size 20 AWG (0.8mm).
+5.5Vdc
DATA
VOICE
GND