User Manual

2
Why choose Agilent oscilloscopes for your toughest high-speed measurement
challenges?
As an engineer you’re no stranger to tough challenges that
help you exceed your customer’s needs and expectations.
Deploying your next design successfully is even more
difficult with today’s high speed technologies. Data signal
eye openings become smaller and measurement error from
your oscilloscope becomes less tolerable. Agilent is
committed to providing the best measurement solutions
for these critical challenges.
Model Real-time
bandwidth on 4 ch
Maximum sampling
rate on 4 ch
Standard memory Maximum
memory
Noise floor at
50 mV/div
91304A 13 GHz 40 GSa/s on 4 ch 20 Mpts on 4 ch 1 Gpts on 4 ch 1.73 mVrms
91204A 12 GHz 40 GSa/s on 4 ch 20 Mpts on 4 ch 1 Gpts on 4 ch 1.45 mVrms
90804A 8 GHz 40 GSa/s on 4 ch 20 Mpts on 4 ch 1 Gpts on 4 ch 1.15 mVrms
90604A 6 GHz 20 GSa/s on 4 ch* 20 Mpts on 4 ch 1 Gpts on 4 ch 0.98 mVrms
90404A 4 GHz 20 GSa/s on 4 ch* 20 Mpts on 4 ch 1 Gpts on 4 ch 0.79 mVrms
90254A 2.5 GHz 20 GSa/s on 4 ch* 20 Mpts on 4 ch 1 Gpts on 4 ch 0.64 mVrms
*DSA model numbers come standard with 50 Mpts of memory on 4 ch.
Sampling Rate 20 Mpts of
memory
50 Mpts of
memory
100 Mpts of
memory
200 Mpts of
memory
500 Mpts of
memory
1 Gpts of
memory
40 GSa/s 500 µs 1.25 ms 2.5 ms 5.0 ms 12.5 ms 25.0 ms
20 GSa/s 1 ms 2.5 ms 5.0 ms 10.0 ms 25.0 ms 50.0 ms
90000A Series Infiniium oscilloscopes
How much time span can I capture?
Note: time span capture = memory depth x 1/ sampling rate
The Agilent Infiniium 90000 Series oscilloscopes are
engineered to give you unmatched real-time measurement
accuracy so you can:
1. Use your jitter budget in your design, not on your
oscilloscope.
2. Pass today’s demanding compliance tests more
quickly.
3. Debug your toughest designs with confidence.