User Manual
47
Measurement Throughput Summary
Typical Cycle Time for Measurement Completion
Cycle Time vs. IF Bandwidth
Cycle Time vs. Number of Points
Data Transfer Time
Cycle time Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with DISPLAY:ENABLE
OFF. Add 21 ms for display on. Data for one trace (S
11
) measurement.
Table 36a. Cycle Time (ms) for Measurement Completion, All Options - Typical
Sweep
Range
IF
Bandwidth
Number of Points
201
401
1601
16001
7 GHz to
8 GHz
600 kHz
Uncorrected
6
6.3
9.6
56
2-Port cal
20
21
28
134
10 kHz
Uncorrected
36
53
200
1945
2-Port cal
80
115
405
3900
1 kHz
Uncorrected
227
444
1740
17000
2-Port cal
460
900
3484
34000
9 GHz to
10 GHz
600 kHz
Uncorrected
6
6.3
9.6
56
2-Port cal
20
21
28
134
10 kHz
Uncorrected
36
53
200
1945
2-Port cal
80
115
405
3900
1 kHz
Uncorrected
227
444
1740
17000
2-Port cal
460
900
3484
34000
10 GHz to
13.5 GHz
600 kHz
Uncorrected
26
33
54
85
2-Port cal
62
77
121
190
10 kHz
Uncorrected
70
118
273
1958
2-Port cal
149
245
553
3922
1 kHz
Uncorrected
236
459
1780
17300
2-Port cal
400
926
3565
34600
13.5 GHz
to 20 GHz
600 kHz
Uncorrected
26
33
54
85
2-Port cal
62
77
121
190
10 kHz
Uncorrected
70
118
273
1958
2-Port cal
149
245
553
3922
1 kHz
Uncorrected
236
459
1780
17300
2-Port cal
400
926
3565
34600