Instruction Manual
Spectrum Master™ MS2712E and MS2713E Specifications
Spectrum Analyzer (continued)
Spectral Purity
SSB Phase Noise @ 1 GHz -100 dBc/Hz, -110 dBc/Hz typical @ 10 kHz offset
-105 dBc/Hz, -112 dBc/Hz typical @ 100 kHz offset
-115 dBc/Hz, -121 dBc/Hz typical @ 1 MHz offset
Amplitude Ranges
Dynamic Range
Measurement Range DANL to +26 dBm
Display Range 1 to 15 dB/div in 1 dB steps, ten divisions displayed
Reference Level Range
Attenuator Range 0 to 55 dB in 5 dB steps
Amplitude Units
Amplitude Accuracy
100 kHz to 4.0 GHz ±1.25 dB, ±0.5 dB typical
>4.0 GHz to 6 GHz ±1.50 dB, ±0.5 dB typical
Displayed Average Noise Level (DANL)
Preamp Off Preamp On
(Reference level -20 dBm) (Reference level -50 dBm)
(RBW Normalized to 1 Hz, 0 dB attenuation) Maximum Typical Maximum Typical
10 MHz to 2.4 GHz -141 dBm -146 dBm -162 dBm
>2.4 GHz to 4 GHz -141 dBm -154 dBm -159 dBm
>4 GHz to 5 GHz -150 dBm -155 dBm
> 5 GHz to 6 GHz -126 dBm -150 dBm
(RBW = 10 Hz, 0 dB attenuation)
10 MHz to 2.4 GHz -152 dBm
>2.4 GHz to 4 GHz -144 dBm -149 dBm
>4 GHz to 5 GHz -124 dBm -140 dBm -145 dBm
> 5 GHz to 6 GHz -116 dBm -121 dBm -140 dBm
Spurs
Residual Spurious <-90 dBm (RF input terminated, 0 dB input attenuation, > 10 MHz)
Input-Related Spurious
Exceptions, typical
Third-Order Intercept (TOI)
Preamp Off (-20 dBm tones 100 kHz apart, 10 dB attenuation)
+16 dBm
2400 MHz +20 dBm
200-2200 MHz +25 dBm, typical
>2.2 GHz to 5.0 GHz
>5.0 GHz to 6.0 GHz
Second Harmonic Distortion
50 MHz -56 dBc
>50 MHz to 200 MHz -60 dBc, typical
VSWR
2:1, typical