Spectrum Master™ MS2712E and MS2713E Specifications
Power Meter (Option 0029)
Frequency Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #, Full Band
Amplitude Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average Acquisition Fast/Med/Slow, # of Running Averages
Limits Limit On/Off, Limit Upper/Lower
Frequency Range 
Span 1 kHz to 100 MHz
Display Range 
Measurement Range -120 dBm to +26 dBm
Offset Range 0 to +100 dB
VSWR 2:1 typical
Maximum Power +26 dBm without attenuator
Accuracy Same as Spectrum Analyzer
Application Options 
High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))
Amplitude Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average # of Running Averages, Max Hold
Zero/Cal Zero On/Off, Cal Factor (Center Frequency, Signal Standard)
Limits Limit On/Off, Limit Upper/Lower
Power Sensor Model PSN50 MA24104A MA24106A MA24108A MA24118A
Description High Accuracy
RF Power Sensor
Inline High
Power Sensor
High Accuracy
RF Power Sensor
Microwave USB
Power Sensor
Microwave USB
Power Sensor
Frequency Range 50 MHz to 6 GHz 600 MHz to 4 GHz 50 MHz to 6 GHz  
Connector     
Dynamic Range 
(.001 to 100 mW)

(2 mW to 150 W)

(0.1 µW to 200 mW)
-40 to +20 dBm
(0.1 µW to 100 mW)
-40 to +20 dBm
(0.1 µW to 100 mW)
VBW 100 Hz 100 Hz 100 Hz 50 kHz 50 kHz
Measurand True-RMS True-RMS True-RMS True-RMS,
Slot Power,
Burst Average
Power
True-RMS,
Slot Power,
Burst Average
Power
Measurement Uncertainty ±0.16 dB
1

2
±0.16 dB
1


Datasheet

11410-00414  11410-00424 11410-00504 11410-00504
Notes: 1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than -20 dBm with
zero mismatch errors.
2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.
Measurement results referenced to the input side of the sensor.
