Technical Data Sheet VNA Master ™ Handheld Vector Network Analyzer + Spectrum Analyzer MS2026C MS2027C MS2028C 5 kHz to 6 GHz 5 kHz to 15 GHz 5 kHz to 20 GHz Vector Network Analyzer MS2036C MS2037C MS2038C 5 kHz to 6 GHz 9 kHz to 9 GHz 5 kHz to 15 GHz 5 kHz to 20 GHz Vector Network Analyzer 9 kHz to 15 GHz 9 kHz to 20 GHz Spectrum Analyzer The Ultimate Handheld Vector Network + Spectrum Analyzer for Cable, Antenna, and Signal Analysis Anytime, Anywhere Introduction High Performance Handheld S-Paramet
Performance and Functional Highlights Vector Network Analyzer • Broadband coverage of 5 kHz to up to 20 GHz • True 2-path, 2-port Vector Network Analyzer • Ultimate accuracy with 12-term error correction • User-defined Quad Display for viewing all 4 S-Parameters • Arbitrary data points up to 4001 • IF Bandwidth selections of 10 Hz to 100 kHz • Directivity: > 42 dB 5 kHz - 5 GHz (all models) > 36 dB 5 GHz - 15 GHz (MS2027C/37C) > 32 dB 15 GHz - 20 GHz (MS2028C/38C) • Transmission Dynamic Range: > 100 dB 2 MH
VNA Master Functional Specifications Definitions All specifications and characteristics apply under the following conditions, unless otherwise stated: • After 30 minutes of warm-up time, when the instrument is in VNA Mode and left in the ON state. • Temperature range is 23 ºC ± 5 ºC. • All specifications apply when using internal reference. • All specifications subject to change without notice. Please visit www.anritsu.com for most current data sheet.
VNA Performance Capabilities (MS202x/3xC) Measurement Parameters S11, S21, S22, S12, Sd1d1, Sc1c1, Sd1c1, Sc1d1 Number of Traces Four: TR1, TR2, TR3, TR4 Trace Format Single, Dual, Tri, Quad. When used with Number of Traces, overlays are possible including a Single Format with Four trace overlays.
Uncertainty Curves for Round-Trip Cable Loss Measurements (1-Port) Round-trip cable loss measurements are convenient for field personnel testing installed cable or waveguide runs. This one-port technique provides one-way data after twice traversing the cable. The following two sets of uncertainty curves, less than 6 GHz on the left and greater than 6 GHz on the right, present worst-case uncertainty by DUT Match (i.e., Log Mag) when using VNA Master for one-port cable loss measurements.
High Port Power OSLxx50 Calibration Components (N-Connectors) Corrected System Performance and Uncertainties: MS202x/3xC Models with 12-term SOLT calibration including isolation using either OSLN50 & OSLNF50 or OSLK50 & OSLKF50 Calibration Kits Precision calibration standards come in a convenient configuration for field work.
Low Port Power OSLxx50 Calibration Components Corrected System Performance and Uncertainties: MS202x/3xC Models with 12-term SOLT calibration including isolation using either OSLN50 & OSLNF50 or OSLK50 & OSLKF50 Calibration Kits Frequency Range (GHz) Directivity (dB) Frequency Range (GHz) Typical High Port Power (dBm) ≤5 > 42 ≤3 –25 ≤ 15 > 36 ≤6 –25 ≤ 20* > 32 ≤ 20 –15 * N Connector guaranteed to 18 GHz, typical > 18 GHz Measurement Uncertainties The following graphs provide measurement un
High Port Power 3652A Calibration Kit (K-Connector) Corrected System Performance and Uncertainties: MS202x/3xC Models with 12-term SOLT calibration including isolation using 3652A Calibration Kit Frequency Range (GHz) Directivity (dB) * Frequency Range (GHz) Typical High Port Power (dBm) ≤5 > 34 ≤3 +3 ≤ 15 > 34 ≤6 –3 ≤ 20 > 34 ≤ 20 –3 * Directivity spec is limited to 34 dB by the 3652A Calibration Kit, not by the instrument performance.
Low Port Power 3652A Calibration Kit (K-Connector) Corrected System Performance and Uncertainties: MS202x/3xC Models with 12-term SOLT calibration including isolation using 3652A Calibration Kit Frequency Range (GHz) Directivity (dB) * Frequency Range (GHz) Typical Low Port Power (dBm) ≤5 > 34 ≤3 –25 ≤ 15 > 34 ≤6 –25 ≤ 20 > 34 ≤ 20 –25 * Directivity spec is limited to 34 dB by the 3652A Calibration Kit, not by the instrument performance.
Spectrum Analyzer Functional Specifications (MS203xC models only) Frequency Frequency Range 9 kHz to 6/15/20 GHz (usable to 0 Hz), Preamp 100 kHz to 6/15/20 GHz Tuning Resolution 1 Hz Frequency Reference Aging: ± 1.0 ppm/10 years Accuracy: ± 0.3 ppm (25 °C ± 25 °C) + aging External Reference Frequencies 1, 1.2288, 1.544, 2.048, 2.4576, 4.8, 4.9152, 5, 9.8304, 10, 13, 19.
Spectrum Analyzer Functional Specifications (MS203xC models only) (continued) Third-Order Intercept (TOI) (–20 dBm tones 100 kHz apart, –20 dBm Ref level, 0 dB input attenuation, preamp Off) 2.
Measurement Options Specifications Time Domain (Option 0002) (includes Distance Domain Option 0501) The VNA Master can also display the S-parameter measurements in the time or distance domain using lowpass or bandpass processing analysis modes. The broadband frequency coverage coupled with 4001 data points means you can measure discontinuities both near and far with unprecedented clarity for a handheld tool.
Bias Tee (Option 0010) Vector Voltmeter (Option 0015) For tower mounted amplifier tests, the MS20x/3xC series with optional internal bias tees can supply both DC and RF signals on the center conductor of the cable during measurements. For frequency sweeps in excess of 2 MHz, the VNA Master can supply internal voltage control from +12 to +32 V in 0.1 V steps up to 450 mA. To extend battery life, an external power supply can substitute for the internal supply by using the external bias inputs instead.
High Accuracy Power Meter (Option 0019) requires external USB power sensor. Conduct precise measurements of CW and digitally modulated transmitters in the field using this VNA Master software mode with a separately purchased Anritsu USB power sensor. After specifying the center frequency and zeroing the sensor to ensure accuracy at low power levels, the software offers intuitive operation for absolute and relative readouts in dBm or Watts. Option 0019 supports the USB Power Sensors in the following table.
GPS (Option 0031) requires external GPS antenna Built-in GPS provides location information (latitude, longitude, altitude) and Universal Time (UT) information for storage along with trace data so you can later verify that measurements were taken at the right location. The GPS option requires a separately ordered magnet mount GPS antenna (2000-1528-R or 2000-1652-R), which are configured to mount outside on a metallic surface.
VNA Master General Specifications (MS202x/3xC) Setup Parameters Size and Weight System Status (Temperature, Battery Info, S/N, Firmware Ver, IP Address, Options Installed) Self Test, Application Self Test GPS (see Option 0031) System Options Name, Date and Time, Ethernet Configuration, Brightness, Volume Language (English, French, German, Spanish, Chinese, Japanese, Korean, Italian, Russian, User defined) Reset (Factory Defaults, Master Reset, Update Firmware) File Save, Recall, Delete, Directory Man
Ordering Information MS2026C 1 MS2027C1 MS2028C1 MS2036C1 MS2037C1 MS2038C1 MS2026C-0002 MS2027C-0002 MS2028C-0002 MS2036C-0002 MS2037C-0002 MS2038C-0002 MS2026C-0005 MS2027C-0005 MS2028C-0005 – – – MS2026C-0007 MS2027C-0007 MS2028C-0007 MS2036C-0007 MS2037C-0007 MS2038C-0007 Secure Data Operation MS2026C-0010 MS2027C-0010 MS2028C-0010 MS2036C-0010 MS2037C-0010 MS2038C-0010 Built-in Bias-Tee – – MS2028C-0011 – – MS2038C-0011 K(f) Test Port Connectors MS2026C-0015 MS20
Ordering Information (continued) Phase-Stable Test Port Cables, Armored 15NNF50-1.5C 15NN50-1.5C 15NDF50-1.5C 15ND50-1.5C 15NNF50-3.0C 15NN50-3.0C 510-97-R 7/16 DIN(f) to 7/16 DIN(f), DC to 7.5 GHz, 50 Ω 1091-379-R 7/16 DIN(f) to 7/16 DIN(f), DC to 6 GHz, 50 Ω, with Reinforced Grip 510-102-R N(m) to N(m), DC to 11 GHz, 50 Ω, 90 degrees right angle 1.5 m, DC to 6 GHz, N(m) to N(f), 50 Ω 1.5 m, DC to 6 GHz, N(m) to N(m), 50 Ω 1.5 m, DC to 6 GHz, N(m) to 7/16 DIN(f), 50 Ω 1.
Waveguide Calibration Components and WG/Coaxial Adapters Recommended waveguide calibration procedure requires two offset shorts and a precision load. The waveguide/coax adapter, shown attached to test port #1, adapts the VNA Master test ports to the waveguide under test. Part Number 1/8 Offset Short 3/8 Offset Short Precision Load Coaxial to Universal Waveguide Adapter[1] 23UM70 24UM70 26UM70 23UM84 24UM84 26UM84 23UM100 24UM100 23UM120 Frequency Range Waveguide Type 35UM70N 5.85 to 8.
The Master Users Group is an organization dedicated to providing training, technical support, networking opportunities and links to Master product development teams. As a member you will receive the Insite Quarterly Newsletter with user stories, measurement tips, new product news and more. Visit us to register today: www.anritsu.com/MUG To receive a quote to purchase a product or order accessories visit our online ordering site: www.ShopAnritsu.