Owner's manual

858.458.0223
sales@ProgrammablePower.com 269
3091LD : Specications
© 2009 AMETEK Programmable Power All rights reserved. AMETEK Programmable Power is the trademark of AMETEK Inc., registered in the U.S. and other countries. Elgar, Sorensen, California Instruments, and Power Ten are
trademarks of AMETEK Inc., registered in the U.S.
Ratings
Power 3000 W @ 0-37° C; 2400 W @ 38-50° C
Current 30 ARMS
Voltage 50 to 350 VRMS
Frequency 45 to 440 Hz
Max. Peak Power 13 kW (up to 20% duty cycle)
Max. Peak/Surge Current 90 APEAK Surge: 300 APEAK for 50 ms
Max. Peak Voltage 500 VPEAK
Operating Modes
Constant Current Range: 0 to 30 ARMS; Accuracy: 0.2 % of full scale; Resolution: 0.05 % of full scale; Min. Set Current: Voltage / Maximum Set Resistance
Constant Voltage Range: 50 to 350 VRMS; Accuracy: 0.2 % of full scale; Resolution: 0.05 % of full scale
Constant Resistance Range: 2.5 to 100 Ohms, 100 to 1000 Ohms; Accuracy: 1% FS, 5% FS; Resolution: 0.05 % FS; Max. Set Resistance: 1 / (Freq.*1.3e-5)
Constant Power Range: 3000 W @ 0-37° C,2400 W @ 38-50° C; Accuracy: 0.5% of full scale; Resolution: 0.1% of full scale
Crest Factor Range: 1.4142 to 4.0, limited to 90 Apeak; Accuracy: 1 % of full scale; Resolution: 0.1 % of full scale
Power Factor Range: 0 to 1 lead or lag limited by Crest factor settings; Accuracy: 1 % of full scale; Resolution: 0.1 % of full scale
Shourt Circuit Mode Max. Surge current: 300 APEAK, up to 50 msec; Max. Cont. current: 30 ARMS; Max. Voltage drop: 2.5 VRMS
Measurements
Frequency Range: 45 to 440 Hz; Accuracy: 0.1% FS; Resolution: 0.05% FS
Voltage Range: 50 to 350 VRMS; Accuracy: 0.1% FS; Resolution: 0.05% FS
Peak Voltage Range: 50 to 500 V; Accuracy: 0.5% FS; Resolution: 0.1% FS
Current Range: 0 to 30 ARMS; Accuracy: 0.2% FS; Resolution: 0.1% FS
Peak Current Range: 0 to 90A; Accuracy: 0.5% FS; Resolution: 0.1% FS
Crest Factor Range: 1,4142 to 4.0; Accuracy: 0.5% FS; Resolution: 0.1% FS
Apparent Power Range: 0 to 3000 VA; Accuracy: 0.5% FS; Resolution: 0.1% FS
True Power Range:0 to 3000 W; Accuracy: 0.5% FS; Resolution: 0.1% FS
Peak Power Range: 0 to 45,000 W; Accuracy: 1% FS; Resolution: 0.1% FS
Reactive Power Range: 0 to 3000 VA; Accuracy: 0.5% FS; Resolution: 0.1% FS
Power Factor Range: 0 to 1; Accuracy: 0.5% FS; Resolution: 0.1% FS
Resistance Range: 2.5 to 100 Ohms,100 to 1000 Ohms; Accuracy: 1% FS, 5% FS; Resolution: 0.05% FS
Protection
Over Current Limited by input - Circuit Breaker and Set Maximum Current Limit in software.
Over Voltage Output protected for voltage transients over 500 V in hardware.
Over Power Power limited at maximum average and peak rated power in hardware. Limited to Set Maximum Power Limit in software.
Over Temperature Monitors heat sink temperature.
Mechanical Specications
Dimensions H: 8.75” (222 mm), W: 16.88” (429 mm), D: 25“ (635 mm)
Weight 74 lbs / 34 Kg.
Remote Sensing Max. 2V drop between sense and load lines
Isolation 1000 V between input and chassis ground
Audible Noise (@1meter) Audible Noise (@ 1meter): Low power: 49.0 dBA; Full power: 69.5 dBA
Control Power Input 115 Vac or 230 Vac (model -230) ± 10%, 47 to 63Hz
Operating Temperature 3000 W @ 0-37° C, 2400 W @ 39-50° C
Interfaces
GPIB Standard
Note: Specications are subject to change without notice. Specications are warranted over an ambient temperature range of 25°± 5° C. Unless otherwise noted, specications are per phase for a sinewave with a resistive
load and apply after a 30 minute warm-up period. For three phase congurations, all specications are for L-N. Phase angle specications are valid under balanced load conditions only.
3000–12000 VA