User guide
14 TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25
FLUXUS® G601 Technical Specification
Step 3
min. medium pressure
Example
Step 4
for the characters 4
to 11 of the transducer order code (ambient temperature, explosion protection, connection
system, extension cable) see page 15
Step 5
for the technical data of the selected transducer see page 16 et seqq.
Lamb wave transducer shear wave transducer
transducer
order code
medium pressure
1
[psi] transducer
order code
medium pressure
1
[psi]
metal pipe plastic pipe metal pipe plastic pipe
min. min. extended min. min. min. extended min.
GLG 218 145 15 GSG 435 290 15
GLH 218 145 15 GSK 435 290 15
GLK 218 (d > 4.7 in)
145 (d < 4.7 in)
145 (d > 4.7 in)
73 (d < 4.7 in)
15 GSM 435 290 15
GLM 145 (d > 2.4 in)
73 (d < 2.4 in)
- 15 GSP 435 290 15
GLP 145 (d > 1.4 in)
73 (d < 1.4 in)
- 15 GSQ 435 290 15
GLQ 145 (d > 0.59 in)
73 (d < 0.59 in)
-15
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
d = inner pipe diameter
step
1 pipe wall thickness in 0.47 0.47 0.47 1.2
selected transducer GLG or GLH GLG or GLH GLG or GLH GS
2 inner pipe diameter in 31.5 23.6 31.5 11.8
max. flow velocity ft/s 49 49 98 49
selected transducer GLG GLG or GLH values not in the range
of the characteristics,
but by using direct
mode, the inner pipe
diameter in the cha-
racteristics is doubled:
GLG
GSK
3 min. medium pressure psi 247 247 247 508
selected transducer GLG GLG or GLH
influence of acoustic
noise is reduced with
increased transducer
frequency, thus
recommended:
GLH
GLG GSK