18 TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25
FLUXUS® G601 Technical Specification
Shear Wave Transducers (extended temperature range)
technical type GDM1EZ7 GDP1EZ7 GDQ1EZ7
order code GSM-ENNNL GSP-ENNNL GSQ-ENNNL
transducer frequency MHz 1 2 4
medium pressure
1
min. extended psi metal pipe: 290 metal pipe: 290 metal pipe: 290
min. psi metal pipe: 435
plastic pipe: 15
metal pipe: 435
plastic pipe: 15
metal pipe: 435
plastic pipe: 15
inner pipe diameter d
2
min. extended in 1.2 0.59 0.24
min. recommended in 1.6 0.79 0.39
max. recommended in 3.1 1.6 0.79
max. extended in 4.7 2.4 1.2
pipe wall thickness
min. in 0.1 0.06 0.04
max.in---
material
housing stainless steel 304 stainless steel 304 stainless steel 304
contact surface Sintimid Sintimid Sintimid
degree of protection NEMA 4 NEMA 4 NEMA 4
transducer cable
type 1699 1699 1699
length ft 13 13 9
dimensions
length l in 2.36 2.36 1.67
width b in 1.18 1.18 0.71
height h in 1.32 1.32 0.85
dimensional drawing
ambient temperature
min. °F -22 -22 -22
max. °F +392 +392 +392
temperature
compensation
xxx
remark on request
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
2
shear wave transducer:
typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 49 ft/s
h
b
l
h
b
l
h
b
l