User guide
TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25 19
Technical Specification FLUXUS® G601
Lamb Wave Transducers
technical type GRG1NC3 GRH1NC3 GRK1NC3
order code GLG-NNNNL GLH-NNNNL GLK-NNNNL
transducer frequency MHz 0.2 0.3 0.5
medium pressure
1
min. extended psi metal pipe: 145 metal pipe: 145 metal pipe:
145 (d > 4.7 in)
73 (d < 4.7 in)
min. psi metal pipe: 218
plastic pipe: 15
metal pipe: 218
plastic pipe: 15
metal pipe:
218 (d > 4.7 in)
145 (d < 4.7 in)
plastic pipe: 15
inner pipe diameter d
2
min. extended in 7.5 4.7 2.4
min. recommended in 8.7 5.5 3.1
max. recommended in 35.4 23.6 11.8
max. extended in 63 39.4 19.7
pipe wall thickness
min. in 0.43 0.28 0.16
max. in 0.91 0.59 0.35
material
housing PPSU with stainless steel
cap 304
PPSU with stainless steel
cap 304
PPSU with stainless steel
cap 304
contact surface PPSU PPSU PPSU
degree of protection NEMA 4 NEMA 4 NEMA 4
transducer cable
type 1699 1699 1699
length ft 16 16 16
dimensions
length l in 5.06 5.06 5.06
width b in 2.01 2.01 2.01
height h in 2.66 2.66 2.66
dimensional drawing
ambient temperature
min. °F -40 -40 -40
max. °F +338 +338 +338
temperature
compensation
xxx
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
2
Lamb wave transducer:
typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended: in reflect arrangement and for a flow velocity of 49 ft/s
pipe diameter max. extended: in diagonal arrangement and for a flow velocity of 82 ft/s
l
h
b
l
h
b
l
h
b