Lamb Wave Transducers
technical type GRM1NC3 GRP1NC3 GRQ1NC3
order code GLM-NNNNL GLP-NNNNL GLQ-NNNNL
transducer frequency MHz 1 2 4
medium pressure
1
min. extended psi - - -
min. psi metal pipe:
145 (d > 2.4 in)
73 (d < 2.4 in)
plastic pipe: 15
metal pipe:
145 (d > 1.4 in)
73 (d < 1.4 in)
plastic pipe: 15
metal pipe:
145 (d > 0.59 in)
73 (d < 0.59 in)
plastic pipe: 15
inner pipe diameter d
2
min. extended in 1.2 0.59 0.28
min. recommended in 1.6 0.79 0.39
max. recommended in 3.5 2 0.87
max. extended in 5.9 2.8 1.4
pipe wall thickness
min. in 0.08 0.04 0.02
max. in 0.2 0.12 0.04
material
housing PPSU with stainless steel
cap 304
PPSU with stainless steel
cap 304
PPSU with stainless steel
cap 304
contact surface PPSU PPSU PPSU
degree of protection NEMA 4 NEMA 4 NEMA 4
transducer cable
type 1699 1699 1699
length ft 13 13 9
dimensions
length l in 2.91 2.91 1.65
width b in 1.26 1.26 0.87
height h in 1.59 1.59 1
dimensional drawing
ambient temperature
min. °F -40 -40 -40
max. °F +338 +338 +338
temperature
compensation
xxx
remark on request
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
2
Lamb wave transducer:
typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended: in reflect arrangement and for a flow velocity of 49 ft/s
pipe diameter max. extended: in diagonal arrangement and for a flow velocity of 82 ft/s
l
h
b
l
h
b
l
h
b