User guide

26 TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25
FLUXUS® G601 Technical Specification
Clamp-on Temperature Probe (optional)
Technical Data
Connection
Temperature Probe
Connector
Cable
technical type PT13N PT13N PT13F PT13F
order code 670413-1 670412-1 670413-2 670412-2
design short response time
type Pt1000 2x Pt1000 matched
according to EN 1434-1
Pt1000 2x Pt1000 matched
according to EN 1434-1
connection 4-wire 4-wire
measuring range °F -22 to +482 -58 to +482
accuracy T
±(0.27 °F + 2
.
10
-3
.
(|T [°F]| - 32 °F)), class A ±(0.27 °F + 2
.
10
-3
.
(|T [°F]| - 32 °F)), class A
accuracy T- 0.1 K, (3 K < T < 6 K),
more corresponding to
EN 1434-1
- 0.1 K, (3 K < T < 6 K),
more corresponding to
EN 1434-1
response time s 50 8
housing aluminum PEEK, stainless steel 304, copper
degree of protection NEMA 4 NEMA 4
weight (without con-
nector)
lb 0.6 1.1 0.7 1.4
fixation clamp-on clamp-on
accessories
thermal conductivity
paste 392 °F
xx
thermal conductivity
foil 482 °F
xx
plastic protection plate,
insulation foam
-x
dimensions
length l in 0.59 0.55
width b in 0.59 1.18
height h in 0.79 1.06
dimensional drawing A B
pin cable of temperature probe extension cable
1 white/blue blue
2red/blue gray
3, 4, 5 not connected
6red red
7white white
8 not connected
cable of temperature probe extension cable
type 4 x 0.25 mm² black or white LIYCY 8 x 0.14 mm² gray
standard length ft 9 16/32/82
max. length ft - 656
cable jacket PTFE PVC
h
l
extension cable
h
l
A
B
red/blue red white/blue white
8
1
2
3
4
5
6
7