User guide
6 TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25
FLUXUS® G601 Technical Specification
Flow Transmitter
Technical Data
FLUXUS G601
design portable
measurement
measurement principle transit time difference correlation principle
flow velocity 0.03 to 115 ft/s, depending on pipe diameter
repeatability 0.15 % of reading ±0.03 ft/s
medium all acoustically conductive gases,
e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propane
temperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
accuracy
volumetric flow rate ± 1 to 3 % of reading ±0.03 ft/s depending on application
± 0.5 % of reading ±0.03 ft/s with field calibration
flow transmitter
power supply 100 to 240 V/50 to 60 Hz (power supply unit),
10.5 to 15 V DC (socket at transmitter),
integrated battery
battery Li-Ion, 7.2 V/4.5 Ah
operating time (without outputs, inputs and backlight): > 14 h
power consumption < 6 W
number of flow measuring
channels
2
signal attenuation 0 to 100 s, adjustable
measuring cycle (1 channel) 100 to 1000 Hz
response time 1 s (1 channel), option: 70 ms
housing material PA, TPE, AutoTex, stainless steel
degree of protection NEMA 4
dimensions see dimensional drawing
weight 4.2 lb
fixation QuickFix pipe mounting fixture
ambient temperature 14 to 140 °F
display 2 x 16 characters, dot matrix, backlight
menu language English, German, French, Dutch, Spanish
measuring functions
physical quantities operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocity
totalizer volume, mass
calculation functions average, difference, sum
diagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times
data logger
loggable values all physical quantities, totalized values and diagnostic values
capacity > 100 000 measured values