User guide
TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25 7
Technical Specification FLUXUS® G601
communication
interface RS232/USB
serial data kit
software (all Windows™
versions)
- FluxData: download of measurement data, graphical presentation,
conversion to other formats (e.g. for Excel™)
- FluxKoef: creating medium data sets
- FluxSubstanceLoader: upload of medium data sets
сable RS232
adapter RS232 - USB
transport case
dimensions 19.7 x 15.7 x 7.5 in
outputs
The outputs are galvanically isolated from the transmitter.
number see standard scope of supply on page 9, max. on request
accessories output adapter (if number of outputs > 4)
current output
range 0/4 to 20 mA
accuracy 0.1 % of reading ±15 μA
active output R
ext
< 200 Ω
passive output U
ext
= 4 to 16 V, depending on R
ext
R
ext
< 500 Ω
frequency output
range 0 to 5 kHz
open collector 24 V/4 mA
binary output
optorelay 26 V/100 mA
binary output as alarm output
- functions limit, change of flow direction or error
binary output as pulse output
- pulse value 0.01 to 1000 units
- pulse width 1 to 1000 ms
inputs
The inputs are galvanically isolated from the transmitter.
number see standard scope of supply on page 9, max. 4
accessories input adapter (if number of inputs > 2)
temperature input
type Pt100/Pt1000
connection 4-wire
range -238 to +1040 °F
resolution 0.01 K
accuracy ±0.01 % of reading ±0.03 K
current input
accuracy 0.1 % of reading ±10 μA
passive input R
i
= 50 Ω, P
i
< 0.3 W
- range -20 to +20 mA
voltage input
range 0 to 1 V
accuracy 0.1 % of reading ±1 mV
internal resistance R
i
= 1 MΩ
FLUXUS G601