Data Sheet | 09.
CONTENTS Base unit specifications ................................................................................................................................................. 5 Timebase TCXO .....................................................................................................................................................................................5 ® Timebase OCXO – R&S CMU-B11 option ...............................................................................................
Power meter (frequency-selective)....................................................................................................................................................24 Modulation analyzer ..........................................................................................................................................................................25 ® R&S CMU-B85 speech codec option ....................................................................................................
Specifications apply under the following conditions: "Typical values" are designated with the abbreviation "typ." These values are verified during the final test but are not assured by Rohde & Schwarz. "Nominal values" are design parameters that are not assured by Rohde & Schwarz. These values are verified during product development but are not specifically tested during production.
Base unit specifications Timebase TCXO Max. frequency drift Max. aging in temperature range +5 °C to +45 °C –6 ±1 × 10 –6 ±1 × 10 /year Timebase OCXO – R&S®CMU-B11 option Max. frequency drift Max. aging in temperature range +5 °C to +45 °C after 30 days of operation Warm-up time at +25 °C –7 ±1 × 10 –7 ±2 × 10 /year –9 ±5 × 10 /day approx. 5 min Timebase OCXO – R&S®CMU-B12 option Max.
RF generator Frequency range Frequency resolution Frequency uncertainty Frequency settling time Output level range RF 1 RF 2 RF 3 OUT Output level uncertainty RF 1, RF 2 RF 3 OUT Output level uncertainty RF 1, RF 2 RF 3 OUT 100 kHz to 2700 MHz 0.
Generator RF level repeatability typical values after 1 h warm-up time output level ≥ –80 dBm output level < –80 dBm <0.01 dB <0.1 dB 10 MHz to 2000 MHz 2000 MHz to 2200 MHz 2200 MHz to 2700 MHz 10 MHz to 2200 MHz 2200 MHz to 2700 MHz 10 MHz to 2200 MHz 2200 MHz to 2700 MHz <1.2 <1.3 <1.6 <1.2 <1.6 <1.5 <1.
Phase noise Carrier offset single sideband, f < 2.2 GHz 20 kHz to 250 kHz 250 kHz to 400 kHz ≥400 kHz Residual FM f < 2.2 GHz 30 Hz to 15 kHz ITU-T (formerly CCITT) Residual AM f < 2.2 GHz ITU-T (formerly CCITT) <–100 dBc, 1 Hz <–110 dBc, 1 Hz <–118 dBc, 1 Hz <50 Hz, rms <200 Hz, peak <5 Hz, rms <0.
Power meter (frequency-selective) Frequency range 10 MHz to 2700 MHz Frequency resolution 0.
Spectrum analyzer Frequency range Span Frequency resolution Resolution bandwidths Sweep time Display Marker Display line Display scale Level range RF 1 RF 2 RF 4 IN Level uncertainty RF 1, RF 2 RF 4 IN Level uncertainty RF 1, RF 2 RF 4 IN Reference level for full dynamic range depending on resolution bandwidth (RBW) 3 continuous power 4 peak envelope power (PEP) continuous power 4 peak envelope power (PEP) continuous power and PEP up to +47 dBm (50 W) up to +53 dBm (200 W) up to +33 dBm (2 W) up to +3
General data Operating temperature range +5 °C to +45 °C, in line with EN 60068-2-1 and -2 –25 °C to +60 °C, in line with EN 60068-2-1 and -2 80 % relative humidity, in line with EN 60068-2-78 Storage temperature range Humidity +40 °C, non-condensing Electromagnetic compatibility in line with EMC Directive 2004/108/EC, applied standard: EN 61326 (immunity for industrial environment; class B emissions) Electrical safety in line with IEC 61010-1: 2001 (ed. 2), EN 61010-1: 2001 (ed. 2), UL 61010-1 (ed.
Inputs and outputs (rear panel) IF 3 RX CH1 Frequency WCDMA other networks/RF Max. output level Impedance BNC female 7.68 MHz 10.7 MHz 0 dBm 50 Ω Remote control interfaces IEC/IEEE bus Serial interface COM 1, COM 2 IEC 60625-2 (IEEE 488.
GSM specifications – mobile station test RF generator Modulation Frequency range GMSK, B × T = 0.
R&S®CMU-B95 additional RF generator option Modulation GMSK, B × T = 0.
Output level range RF 1 8PSK ® without R&S CMU-U99 ® with R&S CMU-U99 –115 dBm to –76 dBm –103 dBm to –64 dBm –103 dBm to –64 dBm RF 2 Output level range RF 1 GMSK overrange mode ® without R&S CMU-U99 ® with R&S CMU-U99 –110 dBm to –28 dBm –90 dBm to –14 dBm –90 dBm to –14 dBm –70 dBm to +9 dBm RF 2 RF 3 OUT Output level range RF 1 8PSK overrange mode ® without R&S CMU-U99 ® with R&S CMU-U99 RF 2 RF 3 OUT –110 dBm to –32 dBm –90 dBm to –18 dBm –90 dBm to –18 dBm –70 dBm to +5 dBm Output level reso
Modulation analysis Level range RF 1 RF 2 RF 4 IN peak envelope power (PEP) see footnote 4 see footnote 4 Inherent phase error GMSK <0.6°, rms <2°, peak Inherent EVM 8PSK ≤1.
Spectrum due to switching Reference level for full dynamic range RF 1 RF 2 RF 4 IN GMSK, RF Attenuation → Low Noise +10 dBm to +47 dBm –4 dBm to +33 dBm –22 dBm to 0 dBm Test method absolute measurement, max. hold over several measurements Filter bandwidth 30 kHz resolution filter, 5 poles Measurement at an offset of ± 400/600/1200/1800 kHz Dynamic range with offset ≥ 1200 kHz >72 dB R&S®CMU-B52 speech codec option Speech decoder output Output impedance Max.
TDMA specifications – mobile station test RF generator Frequency range signaling mode US Cellular PCS (US) 869 MHz to 894 MHz 1930 MHz to 1990 MHz Frequency range non-signaling mode 10 MHz to 2200 MHz Frequency resolution non-signaling mode 1 Hz Frequency uncertainty same as timebase, see base unit specifications Output level range RF 1 RF 2 RF 3 OUT –130 dBm to –32 dBm –130 dBm to –15 dBm –90 dBm to +8 dBm Output level resolution 0.
Power meter (frequency-selective) Level range see base unit specifications Level uncertainty see base unit specifications Power versus time measurement Reference level for full dynamic range RF 1 RF 2 RF 4 IN +4 dBm to +47 dBm –10 dBm to +33 dBm –28 dBm to –6 dBm Dynamic range Filter 100 kHz, rms RF Attenuation Low Noise >74 dB Relative measurement uncertainty result > –40 dB –60 dB ≤ result ≤ –40 dB <0.1 dB <0.
AMPS specifications – mobile station test RF generator Frequency range signaling mode US Cellular 869 MHz to 894 MHz Frequency range non-signaling mode 10 MHz to 2200 MHz Frequency resolution non-signaling mode 1 Hz Frequency uncertainty same as timebase, see base unit specifications Output level range RF 1 RF 2 RF 3 OUT –130 dBm to –27 dBm –130 dBm to –10 dBm –99 dBm to +13 dBm Output level resolution 0.1 dB Output level uncertainty see base unit specifications and add 0.
FM measurement Dynamic range 30 dB below max.
CDMA2000 specifications – mobile station test Standards ® CDMA2000 standards ® CDMA2000 test standards TIA/EIA IS-2000 Rev.
Modulation Dual BPSK, multiple QPSK AWGN Carrier suppression Waveform quality (ρ) factor Code channel level uncertainty Code channel resolution Code channel level range AWGN generator Bandwidth Output level resolution Output level uncertainty Output level range Supported service options Loopback service options Speech service options Test data service option Packet data service option Messaging tele service option Version 09.
RF analyzer Frequency range US/Korean Cellular (band class 0) North American PCS (band class 1) TACS band (band class 2) JTACS band (band class 3) Korean PCS (band class 4) NMT-450 (band class 5) IMT-2000 (band class 6) North American 700 MHz Cellular band (band class 7) 1800 MHz band (band class 8) 900 MHz band (band class 9) secondary 800 MHz band (band class 10) 400 MHz European PAMR (band class 11) 800 MHz PAMR band (band class 12) 2.5 GHz IMT-2000 extension (band class 13) US PCS 1.
Modulation analyzer Level range RF 1 RF 2 RF 4 IN RC1, RC2 (O-QPSK) Waveform quality (ρ) uncertainty Frequency measurement range Frequency measurement uncertainty HPSK, O-QPSK signal f < 2200 MHz f ≥ 2200 MHz f < 2200 MHz f ≥ 2200 MHz f < 2200 MHz f ≥ 2200 MHz –40 dBm to +44 dBm –34 dBm to +44 dBm –54 dBm to +30 dBm –48 dBm to +30 dBm –80 dBm to –9 dBm –74 dBm to –9 dBm waveform quality, error vector magnitude, magnitude error, phase error for ρ 0.
1xEV-DO specifications – access terminal test Standards 1xEV-DO standards 1xEV-DO test standards (access terminal) TIA/EIA IS-856-2 TIA/EIA IS-866-A US/Korean Cellular (band class 0) North American PCS (band class 1) public safety bands (PSB) TACS band (band class 2) JTACS band (band class 3) Korean PCS (band class 4) NMT-450 (band class 5) IMT-2000 (band class 6) North American 700 MHz Cellular band (band class 7) 1800 MHz band (band class 8) 900 MHz band (band class 9) secondary 800 MHz band (band clas
Output level uncertainty RF 1, RF 2 RF 3 OUT Output level uncertainty RF 1, RF 2 RF 3 OUT Modulation Dual BPSK AWGN Carrier suppression Waveform quality (ρ) factor Code channel level uncertainty Code channel resolution Code channel level range AWGN generator Bandwidth Output level resolution Output level uncertainty Output level range Supported applications Test applications Default signaling application Default packet application Version 09.
RF analyzer Frequency range US/Korean Cellular (band class 0) North American PCS (band class 1) public safety bands (PSB) TACS band (band class 2) JTACS band (band class 3) Korean PCS (band class 4) NMT-450 (band class 5) IMT-2000 (band class 6) North American 700 MHz Cellular band (band class 7) 1800 MHz band (band class 8) 900 MHz band (band class 9) secondary 800 MHz band (band class 10) 400 MHz European PAMR (band class 11) 800 MHz PAMR band (band class 12) 2.
Modulation analyzer Level range RF 1 RF 2 RF 4 IN modulated signal f < 2200 MHz f ≥ 2200 MHz f < 2200 MHz f ≥ 2200 MHz f < 2200 MHz f ≥ 2200 MHz –40 dBm to +44 dBm –34 dBm to +44 dBm –54 dBm to +30 dBm –48 dBm to +30 dBm –80 dBm to –9 dBm –74 dBm to –9 dBm Waveform quality (ρ) uncertainty Frequency measurement range Frequency measurement uncertainty for ρ 0.9 to 1 Relative measurement uncertainty result > –33 dB <0.003 –3 kHz to +3 kHz <10 Hz + drift of timebase, see base unit specifications <0.
WCDMA specifications – mobile station (UE) test Standard Symbol rate 3GPP FDD 3.84 MHz Synchronization output 2 Frequency BNC connector REF OUT 2 30.72 MHz/n, n = 1 to 32 RF generator Channels non-signaling mode P-CPICH, P-SCH, S-SCH, P-CCPCH, PICH, DPCH, up to 4 HS-SCCHs, HS-PDSCH, E-AGCH, E-RGCH/E-HICH OCNS R99 OCNS R5 AWGN reference measurement channels (RMC) in line with 3GPP TS 34.
Frequency range non-signaling mode 728 MHz to 768 MHz 869 MHz to 894 MHz 925 MHz to 960 MHz 1475.9 MHz to 1500.
Output level resolution manual mode remote mode 0.1 dB 0.01 dB Output level uncertainty in temperature range +20 °C to +35 °C, no overranging output level ≥ –120 dBm f < 2200 MHz f ≥ 2200 MHz output level ≥ –80 dBm f < 2200 MHz f ≥ 2200 MHz <0.6 dB <0.8 dB in temperature range +5 °C to +45 °C, no overranging output level ≥ –120 dBm f < 2200 MHz f ≥ 2200 MHz output level ≥ –80 dBm f < 2200 MHz f ≥ 2200 MHz <0.9 dB <1.
R&S®CMU-B96 additional RF generator option Path 2 for WCDMA Standard Frequency range 3GPP FDD non-signaling mode 728 MHz to 768 MHz 869 MHz to 894 MHz 925 MHz to 960 MHz 1475.9 MHz to 1500.
Influence on RF interface ® Reduced input level range RF 1 if R&S CMU-B96 is installed continuous input power RF level uncertainty Use of the R&S CMU-B96 in overrange mode may influence all RF signal levels and quality. max. 2 W ® RF analyzer (TX measurements) Frequency range non-signaling mode 698 MHz to 716 MHz 777 MHz to 798 MHz 824 MHz to 849 MHz 880 MHz to 915 MHz 1427.9 MHz to 1452.
Trigger Trigger sources Trigger input Trigger slot delay Trigger delay offset Trigger output signaling mode free run, external, slot, frame, signaling, IF power, TPC, preamble, PRACH-MSG part, compressed mode, change of TFC, HSDPCCH, auto 15-pin D-Sub connector AUX 3, pin 8 external 0 to 14 slots –10239 to + 10239 × 1/4 chip 15-pin D-Sub connector AUX 3, pins 2 to 4 frame, slot, TPC 15-pin D-Sub connector AUX 3, pin 5, preamble, PRACH-MSG part, compressed depending on signaling state and trigger mode, cha
I/Q imbalance Inherent I/Q imbalance Resolution Peak code domain error (PCDE) Inherent PCDE Resolution I/Q analyzer Display <–30 dB 0.01 dB for SF = 4 non-signaling mode QPSK mode and WCDMA mode waveform type zoom rotation measurement length <–40 dB 0.
Spectrum emission mask (SEM application) Measurement filter Display Resolution bandwidth receiver filter in line with standard frequency offset 2.5 MHz to 3.5 MHz 3.5 MHz to 7.5 MHz 7.5 MHz to 8.5 MHz 8.5 MHz to 12.5 MHz 3.84 MHz, RRC, α = 0.22 graphical and numeric values of current, average and max.
Level uncertainty RF 1 RF 2 RF 4 IN in temperature range +5 °C to +45 °C –10 dBm to +47 dBm, rms f < 2200 MHz f ≥ 2200 MHz –44 dBm to –10 dBm, rms f < 2200 MHz f ≥ 2200 MHz –24 dBm to +33 dBm, rms f < 2200 MHz f ≥ 2200 MHz –60 dBm to –24 dBm, rms f < 2200 MHz f ≥ 2200 MHz –24 dBm to 0 dBm, rms f < 2200 MHz f ≥ 2200 MHz –85 dBm to –24 dBm, rms f < 2200 MHz –83 dBm to –24 dBm, rms f ≥ 2200 MHz Level resolution <0.7 dB <0.9 dB <0.9 dB <1.1 dB <0.7 dB <0.9 dB <0.9 dB <1.1 dB <0.7 dB <0.9 dB <0.9 dB <1.
Bluetooth® specifications ® Standards Bluetooth Core Specifications Version 1.
Power meter (frequency-selective) and power versus time Measurement bandwidth filter definition: passband Filter Bandwidth → wide Filter Bandwidth → narrow 2.0 MHz 1.3 MHz Reference level for full dynamic range RF 1 RF 2 RF 4 IN GFSK signal Dynamic range Filter Bandwidth → wide >55 dB, rms Level uncertainty RF 1, RF 2 RF 4 IN in temperature range +20 °C to +35 °C from full scale setting down to –25 dB from full scale setting down to –25 dB <0.7 dB <0.
R&S®CMU-B52 speech codec option Speech decoder output Output impedance Max. output current Full-range output level SPEECH HANDSET OUT 9-pin D-Sub connector <10 Ω 20 mA, peak 1 V, peak Speech coder input Input impedance Full-range input level SPEECH HANDSET IN 9-pin D-Sub connector 100 kΩ 1.4 V, peak 0.1 V, peak Version 09.
R&S®CMU-B17 I/Q/IF interface option I/Q interface Analog I/Q outputs I/Q bandwidth Max. output voltage range IF → I/Q; TX and RX paths, analog I/Q output EMF connector I/Q CH1 0 MHz to 2.5 MHz –1 V to +1 V, peak I 2 Q 2 = 1 V, peak Output impedance I and Q amplitude imbalance Offset voltage Analog I/Q inputs I/Q bandwidth Max. input voltage range Input impedance Carrier suppression Sideband suppression Analog I/Q inputs I/Q bandwidth Max.
IF interface IF inputs, TX path IF level range Standard IF frequencies IF inputs, RX path IF level range Standard IF frequencies IF outputs, TX path IF level range Standard IF frequencies IF outputs, RX path IF level range Standard IF frequencies Version 09.00, October 2008 RF/GSM (GMSK and 8PSK)/ ® TDMA/CDMA2000 WCDMA RF/GSM (GMSK and 8PSK)/ ® TDMA/CDMA2000 WCDMA connector IF3 TX CH1 IN up to –5 dBm, PEP 13.85 MHz 15.36 MHz connector IF3 RX CH1 IN up to +2 dBm, PEP 10.7 MHz 7.
Remarks ® Due to the modulation schemes used Bluetooth and AMPS standards will not be supported. ® ® ® The R&S CMU-B17 and R&S CMU-B73 options use the same mainboard connector of the R&S CMU200. ® ® Therefore, either the R&S CMU-B17 or the R&S CMU-B73 can be ordered for a single instrument. Additional information for GSM: To avoid influences on the fading profile, the following is highly recommended: To set all timeslots to the same level.
R&S®CMU-B41 audio generator/analyzer option AF generator Output impedance <4 Ω Max. output current 20 mA, peak AF sine generator Frequency range Frequency uncertainty Frequency resolution Output level range Output level resolution Output level uncertainty 10 THD+N 10 THD level < 10 mV level ≥ 10 mV level ≥ 1 mV and frequency ≤ 10 kHz level ≥ 100 mV into load ≥ 600 Ω level ≥ 100 mV into load ≥ 600 Ω 20 Hz to 20 kHz same as timebase + half resolution, see base unit specifications 0.
® R&S CMU200 Universal Radio Communication Tester Version 09.
Version 09.
Service you can rely on J J J J J In 70 countries Person-to-person Customized and flexible Quality with a warranty No hidden terms About Rohde & Schwarz Rohde & Schwarz is an independent group of companies specializing in electronics. It is a leading supplier of solutions in the fields of test and measurement, broadcasting, radiomonitoring and radiolocation, as well as secure communications.