Instruction Manual

Version 08.00, June 2011
Rohde & Schwarz R&S
®
FSH4/FSH8 Spectrum Analyzer 13
R&S
®
FSH-K46 CDMA2000
®
BTS pilot channel and EVM measurement
application
R&S
®
FSH-K46E CDMA2000
®
BTS code domain power measurement
application
The specifications below apply to the R&S
®
FSH4 and R&S
®
FSH8. They are based on the data sheet specifications of the R&S
®
FSH4
and R&S
®
FSH8, have not been checked separately and are not verified during instrument calibration. Measurement uncertainties are
indicated as 95 % confidence intervals. The specified level measurement errors do not take into account systematic errors due to
reduced signal-to-noise ratio (SNR).
Measurements R&S
®
FSH-K46 R&S
®
FSH-K46E
Spectrum overview
Result summary
RF channel power
Rho
Carrier frequency error
Active channels
Composite EVM
Peak to average
Pilot channel power (Cd 0)
Sync channel power (Cd 32)
Code domain power
RF channel power
Pilot power
Sync power (rel. to RF ch. pwr./pilot)
Code power (rel. to RF ch. pwr./pilot)
Carrier frequency error
Rho –
Composite EVM
PN offset found
Code domain channel table
Channel type
Walsh code/spreading factor
Symbol rate (ksps)
RC –
Status –
Power absolute (dBm)
Power relative (rel. to RF ch. pwr./pilot)
PN scanner
Detected PN offset
Power per detected PN Offset
All specifications are valid for RC3, one traffic channel, SNR > 30 dB, +15 °C to +35 °C.
Frequency range 15 MHz to 3.0 GHz
Carrier frequency uncertainty, nominal values
Lock range ±10 kHz
Measurement uncertainty SNR > 30 dB,
Δf
ref
= uncertainty of reference frequency
< 10 Hz + Δf
ref
RF channel power
frequency > 15 MHz
preamplifier = OFF –60 dBm < P
RF channel
< 20 dBm
Measurement range
preamplifier = ON –75 dBm < P
RF channel
< 20 dBm
Measurement uncertainty –75 dBm < P
RF channel
< 20 dBm,
ref. level adjusted to RF channel power
< 1 dB, typ. 0.5 dB
PICH power SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm P
RF channel
– 20 dB < P
PICH
< P
RF channel
Measurement uncertainty P
RF channel
– 20 dBm < P
CPICH
< P
RF channel
< 1 dB, typ. 0.5 dB
F-SYNC power SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm P
RF channel
– 20 dB < P
SYNC
< P
RF channel
Measurement uncertainty P
RF channel
– 20 dBm < P
SYNC
< P
RF channel
< 1 dB, typ. 0.5 dB