Instruction Manual

Version 08.00, June 2011
14 Rohde & Schwarz R&S
®
FSH4/FSH8 Spectrum Analyzer
Composite EVM SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm 1.5 % < EVM < 25 %
1.5 % < EVM 10% typ. 2.0 % Measurement uncertainty
10 % < EVM < 25 % typ. 2.5 %
Residual EVM typ. 2.5 %
Rho SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm 0.9 < Rho < 1
0.97 < Rho 1.0 typ. 0.005 Measurement uncertainty
0.90 < Rho 0.97 typ. 0.02
R&S
®
FSH-K47 1xEV-DO
®
BTS pilot channel and EVM measurement
application
The specifications below apply to the R&S
®
FSH4 and R&S
®
FSH8. They are based on the data sheet specifications of the R&S
®
FSH4
and R&S
®
FSH8, have not been checked separately and are not verified during instrument calibration. Measurement uncertainties are
indicated as 95 % confidence intervals. The specified level measurement errors do not take into account systematic errors due to
reduced signal-to-noise ratio (SNR).
Measurements R&S
®
FSH-K47
Spectrum overview
Result summary
RF channel power
Pilot Rho
Carrier frequency error
Traffic activity
Pilot EVM
PN timing (tau)
Peak to average
Pilot power
MAC power
Data power
All specifications are valid for RC3, one traffic channel, SNR > 30 dB, +15 °C to +35 °C.
Frequency range 15 MHz to 3.0 GHz
Carrier frequency uncertainty, nominal values
Lock range ±5 kHz
Measurement uncertainty SNR > 30 dB,
Δf
ref
= uncertainty of reference frequency
< 100 Hz + Δf
ref
RF channel power
frequency > 15 MHz
preamplifier = OFF –60 dBm < P
RF channel
< 20 dBm
Measurement range
preamplifier = ON –75 dBm < P
RF channel
< 20 dBm
Measurement uncertainty –75 dBm < P
RF channel
< 20 dBm,
ref. level adjusted to RF channel power
< 1 dB, typ. 0.5 dB
Pilot power SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm P
RF channel
– 20 dB < P
PICH
< P
RF channel
Measurement uncertainty P
RF channel
– 20 dBm < P
CPICH
< P
RF channel
< 1 dB, typ. 0.5 dB
MAC power SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm P
RF channel
– 20 dB < P
SYNC
< P
RF channel
Measurement uncertainty P
RF channel
– 20 dBm < P
SYNC
< P
RF channel
< 1 dB, typ. 0.5 dB
Data power SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm P
RF channel
– 20 dB < P
SYNC
< P
RF channel
Measurement uncertainty P
RF channel
– 20 dBm < P
SYNC
< P
RF channel
< 1 dB, typ. 0.5 dB
Pilot EVM SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm 1.5 % < EVM < 25 %
1.5 % < EVM 10% typ. 2.0 % Measurement uncertainty
10 % < EVM < 25 % typ. 2.5 %
Residual EVM typ. 2.5 %
Pilot Rho SNR > 30 dB
Measurement range –40 dBm < P
RF channel
< 20 dBm 0.9 < Rho < 1
0.97 < Rho 1.0 typ. 0.005 Measurement uncertainty
0.90 < Rho 0.97 typ. 0.02