Owner's manual
Digital Phosphor Oscilloscopes
TDS7000B Series
Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes
4
Time Base System
TDS7054/TDS7104 TDS7154B/TDS7254B/
TDS7404B/TDS7704B
Time Base Range 200 ps/div to 40 s/div 50 ps to 10 s/div
Time Base Delay Time Range 16 ns to 250 s 5 ns to 250 s
Channel-to-Channel Deskew Range ±25 ns ±75 ns
Delta Time Measurement Accuracy ±((0.06/sample rate) + (15 ppm x reading)) RMS ±((0.06/sample rate) + (2.5 ppm x reading)) RMS
Trigger Jitter (RMS) 8 ps
RMS
(typical) 2 ps
RMS
(typical) (7254B/7154B)
1.5 ps
RMS
(typical) (7404B)
1.2 ps
RMS
(typical) (7704B)
Long Term Sample Rate and Delay ±15 ppm over ≥1 ms interval 2.5 ppm over any ≥100 ms interval
Time Accuracy
Acquisition System
TDS7054 TDS7104 TDS7154B/ TDS7254B/
TDS7404B/ TDS7704B
Real-time Sample Rates
1 channel (max) 5 GS/s 10 GS/s 20 GS/s
2 channels (max) 5 GS/s 5 GS/s 10 GS/s
3-4 channels (max) 2.5 GS/s 2.5 GS/s 5 GS/s
Equivalent Time Sample Rate (max) 250 GS/s 250 GS/s 1 TS/s
Maximum Record Length per Channel 2 Mb (1-CH.), 1 Mb (2-CH.), 500 Kb (4-CH.) 4 Mb (1-CH.), 2 Mb (2-CH.),
with Standard Memory 1 Mb (4-CH.)
With Memory Opt. 2M 8 Mb (1-CH.), 4 Mb (2-CH.),
2 Mb (4-CH.)
With Memory Opt. 3M 16 Mb (1-CH.), 8 Mb (2-CH.),
4 Mb (4-CH.)
With Memory Opt. 4M 32 Mb (1-CH.), 16 Mb (2-CH.),
8 Mb (4-CH.)
With Memory Opt. 5M 64 Mb (1-CH.), 32 Mb (2-CH.),
16 Mb (4-CH.)
Maximum Duration at Highest Real-time Resolution (1-CH)
TDS7054 TDS7104 TDS7154B/TDS7254B/
TDS7404B/TDS7704B
Time Resolution (Single-shot) 200 ps (5 GS/s) 100 ps (10 GS/s) 50 ps (20 GS/s)
Max Duration with Standard Memory 400 µs 200 µs 200 µs
Max Duration with Opt. 2M 1.6 ms 800 µs 400 µs
Max Duration with Opt. 3M 3.2 ms 1.6 ms 800 µs
Max Duration with Opt. 4M 1.6 ms
Max Duration with Opt. 5M 3.2 ms