THE MODULAR SOLUTION FOR 6 KV APPLICATIONS
NSG 3060
691-130B August 2009
Parameter Value
Instrument supply: 230/115 VAC
Decoupling attenuation: Remanent pulse 15% max.
Mains side crosstalk 15% max.
Mains decoupling: 1.5 mH 0% + 35%
Connections: Pulse input from generator
Cable connector for EUT supply
Power inlet for CDN
EUT supply: 1-phase
EUT VAC: 24 to 270 Vrms, 50/60 Hz (phase - neutral), 400 Hz max.
EUT VDC: 0 to 270 VDC
EUT current 1 x 16 Arms continuous (temperature controlled)
1 x 25 Arms for 30 min
EFT (Burst) Standard coupling all lines to ref ground (GND)
IEC/EN 61000-4-4 and ANSI (IEEE) C62.41
L, N, PE ref GND
Any lines and combinations to ref GND:
L ref GND
N ref GND
PE ref GND
L, N ref GND
L, PE ref GND
N, PE ref GND
Combination wave pulse: IEC/EN 61000-4-5
Line to line (2 Ω)
N L / L PE / N PE
Lines to ground (12 Ω)
L PE / N PE / L, N PE
Combination wave & ring wave: ANSI (IEEE) C62.41
Basic 1 & 2 L, N PE & L N
Supplemetal 1 & 2 N PE & L PE
Diagnostic 1 & 2 N, PE L & L, PE N
Ring wave: IEC/EN 61000-4-12
12/30/200 Ω
N L / L PE / N PE
L PE / N PE / L, N PE
PQT: Dips & drops to phase L
Coupling networks CDN 3061