Datasheet

1
Dec
0
Null
Wait 300ms
Command syntax
Alarm,en,1
Alarm,5.5
Alarm,tol,1
Example Response
Alarm,?
Alarm
Alarm,n
Alarm,en,[1,0]
Alarm,tol,n
Alarm,?
enable / disable alarm
sets alarm
sets alarm tolerance (0 – 60)
alarm set?
Wait 300ms
1
Dec
0
NullASCII
?, a l a rm,5.50,1.0 0,1
if all are enabled
1
Dec
0
Null
Wait 300ms
Enable alarm
1
Dec
0
Null
Wait 300ms
O2 level must fall one percentage point
below set point for alarm to reset.
The alarm pin will = 1 when O2 levels are > alarm set point.
Alarm tolerance sets how far below the set point O2 levels
need to drop before the pin will = 0 again.
300ms
processing delay
Alarm
(5.5%)
Alarm
Alarm on Alarm off
(Alarm set point - tolerance)
(5.5% - 1%)
4.5%
Tolerence
(1%)
O
2
43
Copyright © Atlas Scientific LLC