EZO pH Circuit - Datasheet

0.3% from ideal
0.3% from ideal
Fuming
HCL
0.3% from ideal
0.3% from ideal
Fuming
HCL
8.82
T,19. 5
8.91
Temperature compensation
T,n n = any value; floating point or int
T,? compensated temperature value?
Command syntax
r 0.2
T,19.5
RT,19.5
T,?
Example Response
1
Dec
0
Null
Wait 300ms
Wait 900ms
Wait 300ms
8.91
? T,19. 5
1
1
ASCII
ASCII
Dec
Dec
0
Null
0
Null
300ms
processing delay
RT,n
set temperature compensation and take a reading
*
This is a new command
for firmware V2.12
Default temperature = 25°C
Temperature is always in Celsius
Temperature is not retained if power is cut
57