Data Sheet
Constant ow rate
DC,[ml/min], [min or *]
[maintain this rate], [for this much time]
Command syntax
Maintain a constant ow rate
Dispense 25ml per minute
for 40 minutes
DC,25,40
DC,?
Example Response
After running in continuous mode for
20 days the EZO-PMP
TM
will reset.
300ms
processing delay
1
Dec
0
Null
Wait 300ms
Settings that are retained if power is cut
Calibration
Change I
2
C address
Enable/disable parameters
Hardware switch to UART mode
LED control
Protocol lock
Software switch to UART mode
Find
Sleep mode
Settings that are NOT retained if power is cut
evaporation rate = 1ml/min
ow rate = 1ml/min
[ml/min] = a single number (int or oat) representing the desired ow rate
[min or *] = the number of minutes to run or (*) indenitely
A negative value for ml/min = reverse
DC,? reports maximum possible ow rate
?maxrate,58.5
ASCII
1
Dec
0
Null
Wait 300ms
The maximum ow rate is determinded after calibration.
If the owrate entered is to fast the EZO-PMP
TM
will send and error.
*TOOFAST
*ER
r 0.1
55 Copyright © Atlas Scientic LLC