User Manual

Appendix A: Specifications
All specification applies to 10X probe and All ADS1000 Digital Storage
Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope must
first meet the following conditions:
Ɣ The oscilloscope must have been operating continuously for thirty minutes
within the specified operating temperature.
Ɣ You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5° C.
Ɣ The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted “typical.”
Specifications
Inputs
Input Coupling AC, DC, GND
Input Impedance 1Mȍ±2% || 16Pf±3Pf,50ȍ(ADS1202CE,ADS1302CE
contain this option)
Maximum input voltage
400V ˄DC+AC PK-PKˈ1Mȍ input impedance˅ˈCAT
I,CAT II
Probe attenuator
1Xǃ10X
Probe attenuator Factors
Set
1Xǃ10Xǃ100Xǃ1000X
Vertical System
Vertical Sensitivity
2mV/div - 5V/div(1-2-5 order)ADS1000C/CE
2mV/div - 10V/div(1-2-5 order)ADS1000CA
Channel voltage
offset range
ADS1000C, 2mV–100mV:±2V
200mV - 5V:±40V
ADS1000CA, 2mV-100mV: ±2V
200mV-10V: ±40V
ADS1000CE, 2mV –100mV: ±800mV
123