User Manual
1.3.2 Probe Compensation
As an alternative method to Probe Check, you can manually perform this
adjustment to match your probe to the input channel.
Picture1-7
1. Set the Probe option attenuation in the channel menu to 10X. Set the switch to 10X
on the probe and connect the probe to channel 1 on the oscilloscope. If you use the
probe hook-tip, ensure a proper connection by firmly inserting the tip onto the
probe.
2. Attach the probe tip to the PROBE COMP̚3V connector and the reference lead
to the PROBE COMP Ground connector. Display the channel and then push the
“AUTO” button.
3. Check the shape of the displayed waveform.
Over compensated Compensated correctly under compensated
Picture 1-8
4. If necessary, adjust your probe. Repeat as necessary.
8