Note˖Ɣ The oscilloscope vertical response rolls off slowly above its bandwidth ,
which depending on the model, or 20 MHz when the Bandwidth Limit
option is set to “On”. Therefore, the FFT spectrum can show valid
frequency information higher than the oscilloscope bandwidth. However,
the magnitude information near or above the bandwidth will not be
accurate.
Ɣ If the channel is set to DC couplingˈyou can quickly measure the DC
component of the signal by simply nothing its distance from the ground
symbol.
Ɣ If the channel is set to AC couplingˈthe DC component of the signal is
blocked allowing you to use greater sensitivity to display the AC
component of the signal.
Setting up the CH1ǃCH2 Channels
Each channel has its own separate Menu. The items are set up separately
according to each channel.
1. Set up the channel couple
Take the CH1 for example; the tested signal is a sine wave signal with DC
deflection:
Ɣ Press“CH1”ĺ“Coupling”ĺ“AC”, Set to AC couple mode. It blocks the DC
component of the input signal.
Set to AC
cou
p
lin
g
Picture 2-6
Ɣ Press“CH1”ĺ“Coupling”ĺ“DC”, Set to DC couple mode. Both
AC status
DC and AC component could be obstructed.